Research Article

Trace and Essential Elements Analysis in Cymbopogon citratus (DC.) Stapf Samples by Graphite Furnace-Atomic Absorption Spectroscopy and Its Health Concern

Table 1

AAS Vario-6 Graphite furnace elements instrumental analytical conditions.

ElementsWavelength
(nm)
Slit width (nm)Atomization temp. (0°C)Matrix modifiersInterference wavelength (nm)Detection limit
(µg L−1)

Mg 285.20.81500–16500.0025
Ca 422.71.22350–25000.007
Cr357.90.82100–2200NH4H2PO4Fe 358.1, Nb 358.00.1
Mn279.50.21600–1650Mg(NO3)2
+ Pd(NO3)2
Mg 279.5, Fe 279.5,
Pb 280.2
0.014
Fe248.30.21850–2050Mg(NO3)20.100
Ni 232.00.22100–2300Mg(NO3)20.3
Cu324.80.81800–1900Ni 324.3, Mn 324.9,
Pd 324.3, Ag 324.8,
Eu 324.8
0.19
Zn213.90.81000–1100Pd(NO3)2Cu 213.9, Te 214.3
As 214.4, Fe 213.6,
Fe 213.9
0.002
Mo 313.30.82500–2600Mg(NO3)20.15
As 193.70.02050–2250Pd(NO3)20.4
Cd228.80.8900–1200NH4H2PO4 + Mg(NO3)2As 228.9
Fe 228.8
0.007
Pb217.00.51200–1350 Pd(NO3)2
+ Mg(NO3)2
Cu 216.5, Fe 216.7,
Ni 216.6, Sb 217.6,
Pt 216.5
0.08