Research Article

Performance Evaluation of Malaria Pf/Pv Combo Test Kit at Highly Malaria-Endemic Area, Southern Ethiopia: A Cross-Sectional Study

Table 3

Overall performance characteristics of Care Start™ in comparison with the standard light microscopy, Arba Minch, 2016.

Performance characteristicsCare Start™ malaria

Sensitivity (95% CI)97.44% (91.04%–99.69%)
Specificity (95% CI)93.67% (85.84%–97.91%)
Positive predictive value (95% CI)93.83% (86.18%–97.97%)
Negative predictive value (95% CI)97.37% (90.82%–99.68%)