Research Article

Calculation of Toric Intraocular Lens Power with the Barrett Calculator and Data from Three Keratometers

Table 2

Mean anterior corneal keratometry values obtained using iTrace, IOLMaster, and Pentacam.

iTrace at 3 mmIOLMaster at 2.3 mmPentacam at 3 mm
(n = 101)(n = 101)(Centered on the corneal apex) (n = 101)

Astigmatism magnitude (D)1.86 ± 0.771.95 ± 0.871.83 ± 0.79
Range: 0.61–4.23Range: 0.34–4.50Range: 0.40–3.80
Astigmatism axis (degree)97 ± 6899 ± 6494 ± 47
Range: 0–179Range: 1–179Range: 0–179
Flat keratometry (D)43.69 ± 1.5844.04 ± 1.6144.01 ± 1.58
Range: 39.58–47.74Range: 39.57–47.74Range: 39.40–47.60
Steep keratometry (D)45.55 ± 1.6245.99 ± 1.5745.84 ± 1.60
Range: 41.02–48.77Range: 42.13–49.41Range: 41.10–49.10
Mean keratometry (D)44.62 ± 1.5545.02 ± 1.5344.92 ± 1.54
Range: 40.30–48.26Range: 40.85–48.51Range: 40.25–48.25

D: diopter.