Research Article

Calculation of Toric Intraocular Lens Power with the Barrett Calculator and Data from Three Keratometers

Table 3

Mean difference between anterior corneal keratometry values for each pair of devices.

iTrace-IOLMasteriTrace-PentacamIOLMaster-Pentacam

Astigmatism magnitude (D)−0.10 ± 0.330.03 ± 0.270.12 ± 0.38
 = 0.005 = 0.325 = 0.002
Flat keratometry (D)−0.35 ± 0.27−0.31 ± 0.330.04 ± 0.36
 < 0.001 < 0.001 = 0.310
Steep keratometry (D)−0.45 ± 0.26−0.29 ± 0.300.16 ± 0.24
 < 0.001 < 0.001 < 0.001
Mean keratometry (D)−0.40 ± 0.20−0.30 ± 0.290.10 ± 0.24
 < 0.001 < 0.001 < 0.001

D: diopter.