Broadband Bioimpedance Spectroscopy Based on a Multifrequency Mixed Excitation and Nuttall Windowed FFT Algorithm
Table 4
Experiment results of BIS measurement under the excitation of the MFM signal
.
| Primary harmonic | Frequency (kHz) | Standard s | Measured
| Relative errors (%) | Standard Ps | Measured Pm | Absolute errors (°) |
| | 3.90625 | 918.1501 | 918.1311 | 0.0021 | | | 0.0148 | | 7.8125 | 912.6795 | 912.8680 | 0.0207 | | | 0.0183 | | 15.625 | 891.9151 | 891.8473 | 0.0076 | | | 0.0057 | | 31.25 | 823.2610 | 822.8110 | 0.0547 | | | 0.0495 | | 62.5 | 668.0581 | 667.5680 | 0.0734 | | | 0.0450 | | 125 | 487.1455 | 486.5256 | 0.1273 | | | 0.0881 | | 250 | 381.8698 | 381.0998 | 0.2016 | | | 0.0819 | | 500 | 344.2257 | 343.3310 | 0.2599 | | | 0.0524 | | 1000 | 333.6487 | 332.7088 | 0.2817 | | | 0.0263 |
|
|