Research Article

Research on Tuning Fork Dimension Optimization and Density Calculation Model Based on Viscosity Compensation for Tuning Fork Density Sensor

Table 5

Comparison between simulation results and actual measurement results of resonant frequency.

Density of liquid (g/cm3)Resonant frequency (Hz)The simulation of frequency (Hz)Relative error (%)

0.8453982.12985.650.457
0.8611980.59985.070.511
0.8802979.81984.190.447
0.9111977.57982.080.461
0.9511974.69979.380.567
0.9852972.28976.700.455
1.0762966.32971.080.461
1.1058965.15969.180.417
1.1281963.01967.680.485
1.1661960.22965.230.522
1.2103957.82962.410.479
1.2251956.91961.450.474
1.2453955.50959.660.435
1.2653954.69959.110.463
1.2761953.45958.070.485