Research Article
Research on Tuning Fork Dimension Optimization and Density Calculation Model Based on Viscosity Compensation for Tuning Fork Density Sensor
Table 5
Comparison between simulation results and actual measurement results of resonant frequency.
| Density of liquid (g/cm3) | Resonant frequency (Hz) | The simulation of frequency (Hz) | Relative error (%) |
| 0.8453 | 982.12 | 985.65 | 0.457 | 0.8611 | 980.59 | 985.07 | 0.511 | 0.8802 | 979.81 | 984.19 | 0.447 | 0.9111 | 977.57 | 982.08 | 0.461 | 0.9511 | 974.69 | 979.38 | 0.567 | 0.9852 | 972.28 | 976.70 | 0.455 | 1.0762 | 966.32 | 971.08 | 0.461 | 1.1058 | 965.15 | 969.18 | 0.417 | 1.1281 | 963.01 | 967.68 | 0.485 | 1.1661 | 960.22 | 965.23 | 0.522 | 1.2103 | 957.82 | 962.41 | 0.479 | 1.2251 | 956.91 | 961.45 | 0.474 | 1.2453 | 955.50 | 959.66 | 0.435 | 1.2653 | 954.69 | 959.11 | 0.463 | 1.2761 | 953.45 | 958.07 | 0.485 |
|
|