Abstract

This study evaluates the sensitivity of microelectromechanical system (MEMS) capacitive bending strain sensor with a double layer cantilever designed to meet the requirements of spinal fusion monitoring. The cantilever structure of the sensor consists of two parallel substrate plates which constitute the electrodes, attached to an anchor made of silicon dioxide. The sensor was able to monitor bending strain value ranging from 0 to 1000 με. In order to evaluate the sensitivity of the sensor, parametric study was carried out by varying electrode gap, anchor length, and dielectric coverage between the electrodes. The nominal capacitive strain sensor for various applications has sensitivity ranging from 255 aF/με to 0.0225 pF/με. An increase in the sensitivity was observed on reducing the electrode gap and the anchor length and increasing the dielectric coverage, resulting in a highest sensitivity value of 0.2513 pF/με. It was also observed that dielectric constant has a significant effect on the sensitivity behavior of the sensor.

1. Introduction

MEMS capacitive based sensor has been widely used and has a variety of applications that includes measurement of pressure, gas flow, force, position, displacement, chemical, vacuum, tactile, and RH monitoring. When compared to piezo based sensors, capacitive sensors have several advantages such as high sensitivity, low power consumption, better temperature performance, and less sensitivity to drift [1]. By varying the variable in the capacitance equation the capacitance output of the sensor can be manipulated.

Several types of capacitive sensors have been developed for strain monitoring applications. A capacitive strain sensor for automobile tire strain monitoring, having a comb drive structure, was developed by Matsuzaki and Todoroki [2]. The sensor was located on the flexible substrate. When subjected to strain, the gap between the comb drive electrodes increased causing the difference in values of capacitance. They observed that sensitivity of the sensor varied with design and the highest sensitivity value obtained was at 45.71 aF/με. Suster et al. [3] also developed a MEMS capacitive strain sensor with a comb drive structure for rotating shaft surface strain monitoring. When strain was induced, the area of the overlap electrodes changed significantly leading to change in the capacitance output of the sensor. This sensor exhibited higher sensitivity at 255 aF/με. Aebersold et al. [4] in their work studied the MEMS capacitive bending strain sensor for spinal fusion monitoring. They developed a sensor having a comb drive structure, with electrodes made from heavily doped silicon.

When bending occurred, the gap and also the overlap area of the electrode changed, resulting in the change in capacitance reading. The sensitivity of this sensor was 2.94 fF/με for strain value between 0 to 1000 με. Similarly, Lin et al. [5] developed a MEMS capacitive bending strain sensor for spinal fusion monitoring but having different sensor structure. The sensor had a cantilever beam structure with two layers. When bending strain was induced, the gap between the electrodes changed, which led to the capacitance change for bending strain reading. The sensor exhibited highest sensitivity of 0.02 pF/με between 0 to 1000 με.

A comparative study of different capacitive strain sensors with their sensitivity values has been summarized in Table 1. The comb drive structure can be designed to manipulate the gap, and also the overlap area of the electrodes to induce the capacitance change. It is also the most commonly used structure for sensor. However, the cantilever beam structure designed by Lin et al. [5] showed highest sensitivity when compared to comb drive structure. In this study, the cantilever beam structure design has been utilized owing to its higher sensitivity.

The primary objective of this work was to evaluate the sensitivity of MEMS capacitive bending strain sensor, designed using cantilever beam structure. Equation (1) demonstrates that the capacitance output can be increased by reducing the electrode gap, increasing the overlap area of the electrodes or by using a high dielectric coefficient material between the electrodes. For any sensor, these parameters are of significant importance and are highly dependent on the structural design of the sensor. In case of cantilever beam structure design, actuated due to bending forces, all these parameters can be applied to increase the capacitance output. However, the influence of these parameters on the sensitivity of the sensor to the bending strain is not well established. Therefore this study was carried out to ascertain the relationship between these parameters on the sensitivity of the designed sensor. The present sensitivity enhancement study was accomplished by varying three parameters, namely, electrode gap, anchor length, and the dielectric material. By changing these parameters the capacitance output of the sensor and its sensitivity when bending strain developed can be examined. Simulation using finite element software, CoventorWare 2008, was also carried out and the results are compared with those of analytical output.

2. Sensor Design

The sensor was designed to meet the requirements of spinal fusion monitoring. Spinal fusion is a medical procedure, wherein the lumbar vertebrate causing pain to the patient needs subsequent fixation of titanium rod to facilitate healing [6, 7]. The designed sensor will be located on the titanium rod fixation and monitors the bending strain change during the healing process. Kanayama et al. [8] in their work concluded that bending strain was the main loading anticipated by the titanium rod. Gibson [9] showed that the maximum strain value of the rod was nearly 1000 με. In order to be able to mount on the rod and other wireless circuitry, Aebersold et al. [4] and Lin et al. [5] suggested the maximum dimension of the sensor to be around 10 mm × 2 mm (width × length). Table 2 summarizes the requirement of the sensor for spinal fusion monitoring.

In this work, the sensor designed has three layers. The substrates at the bottom and top layer are the electrodes built from heavily doped silicon for better conductivity. The gap between these electrodes was determined using the anchor made from silicon dioxide. This gap was varied to study the capacitance output and also the sensitivity. The anchor length was also varied as it would affect the opening of the electrodes when bending occurred. Finally, the effect of dielectric material with three different percentage configurations was also determined. The sensor designed had width of 2 mm. The schematic diagram of the sensor in Figure 1 shows the three parameters used.

3. Theory

The sensor actuation can be modeled as a simple cantilever beam bending, which has one fixed end and a suspended free end on the other side as shown in Figure 2.

The capacitance of the two electrode plate can be written as The relationship for a simple beam bending can be determined from the beam bending theory and the flexure formula as [10]

Moment along was given by where is the beam width and is evenly distributed pressure load. Slope, , and deflection, along -axis were given by

The sensor was modelled on the simple cantilever beam. The beam was fixed at one end, and bending occurred at the free end. During bending, the lower electrode of the sensor also bends along with the cantilever beam. Meanwhile, the top electrode of the sensor, fixed along point to , will have a tangent deflection [11]. The top electrode of the sensor would bend in a tangent form starting at point . The mechanism would increase the gap of the electrodes as the bending occurred. The deflection of the sensor model is illustrated in Figure 3.

Tangent deflection equation at point along -axis can be written as where is a constant that can be obtained using a boundary condition [5, 11]. When the sensor model bends, the tangent deflection line at point was equal to the deflection of the lower layer of the sensor, , one has

Using (4) and (5), the constant can be obtained as then is substituted into (5) and gives us

Thus, the difference in deflection or gap between the tangent deflection line at point , , and the lower layer of the sensor model beam along -axis can be written as and can be given by the following equation:

is integrated for and to acquire the average value of the gap along axis that corresponds to pressure load, applied, one has

Solving (2), (3), and (11) gives (12) used for calculating the average electrodes gap for sensor. The value for and depends on the analysis study for sensitivity enhancement. Table 3 gives the typical values used for the analysis.

The total capacitance for electrode gap and anchor length analysis, was given by

For dielectric coverage analysis, the total capacitance was given by

4. Results and Discussions

Simulation was carried out using MEMS conventional finite element software, CoventorWare 2008. The boundary conditions applied for the simulation were exactly similar to the analytical method and hence a similar bending mechanism is expected, with minimum errors and uncertainties. Figure 4 shows the boundary conditions applied in the finite element study. The sensor was placed on the cantilever beam having one end fixed, and a distributed load applied uniformly along the bottom surface of the cantilever beam.

4.1. Electrode Gap (Anchor Thickness)

The effect of varying electrode gap is studied. Based on the sensor design, three different gap values, 1.5 μm, 3 μm, and 5 μm, are selected. The anchor length was maintained constant at 1 mm. Figure 5 shows the results of electrode gap comparison for the analysis study.

The electrode gap of 1.5 μm demonstrated higher capacitance output, followed by the 3 μm and 5 μm gap. Table 4 shows the results for sensitivity, nominal capacitance, and the average and maximum percentage difference between calculation and simulation results.

The lowest gap 1.5 μm exhibits the highest sensitivity between 0 and 1000 με, followed by 3 μm and 5 μm. The sensitivity increased by 129% when the gap was reduced to half of its original length. From Figure 6 it can be inferred that the capacitance change was significant when load was applied between strains 0 to 200 με. In this region the sensor was very sensitive to the strain changes. The variation of sensitivity for different region of strain can be observed in Figure 6.

4.2. Anchor Length

Another parameter that was considered for this study was the length of the anchor. The length of anchor determines the overall electrode gap opening when bending occurred. Longer the anchors, higher the nominal capacitance, but a longer anchor gives a lower capacitance drop when bending occurs. Three different anchor lengths, 0.5 mm, 1.0 mm, and 1.5 mm, were used to study the capacitance output of the sensor. The initial gap of the electrode was set at 1.5 μm. Figure 7 shows the anchor length results when bending occurred.

The results with 1.5 mm of anchor length showed highest capacitance output. This can be attributed to larger dielectric area present. The dielectric area with a dielectric coefficient of 3.9 showed increase in the capacitance output. Despite this, 1.5 mm anchor length displayed lower sensitivity when compared to others. However, 0.5 mm anchor length had lowest capacitance output but showed the highest sensitivity. Table 5 shows the analysis results for various anchor lengths.

The sensitivity of the sensor decreased with increase in the anchor length. When the length of the anchor was small, the movable area of the sensor increased. This led to an increase in the electrode gap opening during bending and resulted in the higher capacitance drop and also the sensitivity of the sensor. These results with capacitance output were similar to that obtained by varying the electrode gap. The change in capacitance decreased when strain was increased; however the sensor showed a greater change in capacitance in the strain range of 0–200 με. Figure 8 shows the sensitivity of the sensor for various strain values.

4.3. Dielectric Coverage

The dielectric material study covering 33%, 66%, and 100% of the electrode area was undertaken. The 100% coverage area gave the highest capacitance output. The 33% and 66% coverage area was located near the opening end of the sensor. While bending, opening end of the sensor would have highest gap change, and therefore by locating the dielectric material in this location, the change in capacitance can be expected to be higher. The anchor length was set at 1 mm and the electrodes gap of 1.5 μm was utilized for this study. Figure 9 shows the results for each of the dielectric covers employed in this study.

The presence of dielectric led to a significant increase in the capacitance output. During bending, the capacitance output dropped from 75% to 82% for a total capacitance change between 0 and 100 με. Most of the capacitance change occurred in this region. The capacitance output between 100 and 1000 με for 66% and 33% of dielectric coverage, was almost identical. Table 6 summarizes the results for dielectric coverage studies.

The sensitivity and capacitance output increased with the dielectric coverage. It was observed that, reducing dielectric material or locating it at the opening end of the sensor beam does not increase the sensitivity of the sensor. Figure 10 shows the sensitivity for all the three dielectric coverage, compared at different strain ranges.

4.4. Simulation Analysis Discussion

Simulation study was carried out involving three parameters in order to study the sensor response to bending strain in terms of capacitance output and sensitivity. Figure 11 shows the nominal capacitance of the sensor when different gap and different anchor length were applied. It is evident from these results that by reducing electrode gap, increasing anchor length and increasing dielectric coverage would in turn lead to increase in capacitance output.

On reducing the electrode gap, a nonlinear increase in the nominal capacitance was observed. The gap represents a significant factor in the capacitance output of the sensor. However, a very small gap can cause the electrodes to touch if the electrostatic force applied was strong enough. Therefore, for a 1.5 μm electrode gap, the highest voltage that can be applied for the capacitance reading was . A higher voltage can lead to sensor deflection which can change the capacitance reading even before the application of bending strain. Figure 12 illustrates the deflection of the sensor and the capacitance change due to electrostatic force. The deflection occurred at the top electrode of the sensor as shown in Figure 13. The deflection of the top layer of the sensor decreased when the voltage decreased. For a voltage of 0.1 V, a very high capacitance value of 117.305 pF was observed as a result of deflection in the reduced gap. However, when the voltage applied was , the change in capacitance was very low and when the voltage applied was , the change in capacitance was almost negligible.

With respect to the anchor length, changes in the nominal capacitance are of linear order. The anchor basically is composed of silicon dioxide, for which the coefficient of dielectric material was 3.9. For a nominal capacitance, increasing the length of the anchor would also increase the dielectric material coverage between the electrodes. So, the nominal capacitance would increase in the same manner for both anchor length analysis and dielectric coverage analysis.

During cantilever beam bending, the capacitance output would follow nonlinear pattern with the strain. Similarly, the capacitance drop was different for other variations of electrode gap, anchor length, and total dielectric coverage on the movable area. The capacitance drop was the result of the change in electrode gap at the movable part of the sensor. When bending increased, the gap also increased, because of which the capacitance output dropped. Thus the capacitance change corresponding to the bending strain can be evaluated using the sensitivity value. The sensitivity of the sensor increased when the electrode gap was reduced. During bending, the gap between the electrodes increased. The electrode gap which was small initially would increase significantly with bending, in comparison to the one with bigger initial electrode gap for the same bending strain value. The greater percentage of gap increment led to increase in the capacitance drop and also the sensitivity of the sensor. This is further demonstrated in Figure 14 which shows the average value of gap percentage increment for different electrodes gap.

In case of anchor length analysis, a completely different sensitivity performance was observed. The sensitivity of the sensor decreased when the anchor length was increased, even though the nominal capacitance was larger for the longer anchor length. This capacitance drop can be attributed to the movable part of the sensor. Despite the fact that longer anchor length gave a higher capacitance output, the movable part of the sensor was shorter. A shorter movable electrode results in smaller capacitance drop when bending occurred and thus reduced the sensitivity of the sensor. Figure 15 shows the comparison for the average percentage drop of capacitance output at the movable part of the sensor electrodes for different anchor length. Anchor length of 0.5 mm showed a greater average percentage drop in capacitance. This can be explained due to the longer length of movable part when bending occurred.

The last sensitivity analysis was that of dielectric coverage. The large portion of dielectric coverage would increase the capacitance output and sensitivity of the sensor. 100% dielectric coverage displayed the maximum sensitivity. However, the 33% and 66% of dielectric coverage did not increase the sensitivity as expected. During the bending process, the electrode gap from the extreme free end of the electrode towards the anchor is reduced to a minimum. Thus, by placing the dielectric material having dielectric constant of 3.9, at the free end of the electrodes and covering 33% or 66%, the capacitance drop when bending occurred was expected to be higher. However, the results showed a different behaviour. From the study, it was found that the sensitivity of 33% and 66% did not increase due to the anchor capacitance. The anchor which was made from silicon dioxide contributed 46.042 pF of capacitance. This capacitance value did not change with bending strain because no gap increment occurred in this area. Thus, the total capacitance which includes the anchor capacitance reduced the total capacitance drop of the sensor and also the sensitivity of 33% and 66% dielectric coverage sensor. However, the average percentage capacitance drop for 33% and 66% which occurred at the movable part of the electrode was higher when compared to 100% coverage. Figure 16 shows the average percentage capacitance drop comparison at the movable electrode area for all dielectric coverage.

The 33% dielectric coverage showed a higher capacitance drop, followed by 66% and 100% coverage at movable electrodes of the sensor. This shows that if the anchor material was to be substituted with a lower dielectric constant, it can reduce the capacitance of the anchor; thereby the sensitivity of 33% coverage can be optimized and a higher value than 100% dielectric coverage was achievable.

The overall sensitivity performance of the sensor indicates that the sensor was very sensitive at the early stages of bending strain development. The electrode gap and anchor length analysis study showed that the sensor had a higher sensitivity value ranging between 0 and 200 με, and the sensitivity remained the same on further increase in the strain beyond 200 με. On the other hand the dielectric coverage analysis study showed that the highest sensitivity was between 0 and 100 με. This sensitivity performance can be attributed to the percentage gap difference when bending occurred. At the beginning of the bending process, the percentage gap increment was higher, allowing a higher capacitance drop. Later as the bending increased further, the percentage gap increment was reduced leading to lower capacitance drop and lower sensitivity. This aspect of percentage gap increment with bending strain can be observed for all the cases in Figure 17.

From the figure it is evident that the 33% and 66% gap increment for air part was lesser because it was located near the anchor which facilitated only a small opening. The initial gap for dielectric area with 33%, 66%, and 100% can be assumed to be 0 μm. Thus, during the start of bending process, the percentage gap increment was very high leading to higher sensitivity. It is for this reason the sensitivity between 0 and 100 με was very high for dielectric coverage sensor.

Figure 18 shows the contours for strain distribution when the sensor was attached to the loaded beam. The strain showed the highest value at the end of anchor region. It is clear from the figure that the stress was concentrated at this region during the bending process. Observations were obtained by Lin et al. [5] who found that the maximum strain before fracture for silicon was around 1800 με. Therefore on further loading anchor region of the sensor would fail due to fracture.

5. Conclusions

Simulation study was carried out to understand the MEMS capacitive bending strain sensor based on double layer cantilever design of behaviour and its performance on application of bending strain. The capacitance output, sensitivity performance evaluation by manipulating the electrodes gap, anchor length, and dielectric coverage were carried out and validated using analytical methods and the results showed that nominal capacitance and sensitivity of the sensor increased when the gap between the electrodes was reduced. In addition, it was found that sensitivity of the sensor increased with a shorter anchor length as a result of higher percentage gap opening during the bending process. The 100% dielectric coverage using silicon dioxide material demonstrated highest sensitivity. Meanwhile the sensitivity of 33% and 66% dielectric coverage can be optimized using dielectric material having lower dielectric constant than the sensor anchor. The overall sensor performance indicated that the sensor can be optimized by reducing the electrode gap, reducing the anchor length, and on application of dielectric material. It was concluded that a very small voltage of volt was preferable for capacitance reading of the sensor as it helped to overcome the electrostatic forces developed. However, while designing any sensor, factors such as fabrication abilities and electrostatic charges which add to the limitation of the sensor performances should also be considered.

Nomenclature

: Relative permittivity of free space
: Dielectric constant
: Overlap area of the electrode
: Gap of the electrode
: The maximum normal stress in the member
: The resultant internal moment
: The moment of inertia of the cross-sectional area computed about the neutral axis
: The farthest point perpendicular to neutral axis
: Normal stress in the member
:Young’s Modulus
:Strain
: Total capacitance
: Capacitance of anchor
: Capacitance of air gap
: Multilayer dielectric capacitance
: Capacitance of air area (with dielectric)
: Capacitance of dielectric area
: Capacitance change between two points of strain
: Capacitance at 0 μɛ
: Capacitance at 1000 μɛ
: Strain change at two points
: Anchor area
: Air gap area (no dielectric)
: Air gap area (with dielectric)
: Dielectric area
: Initial gap (dielectric area)
: Dielectric gap
: Average gap of the electrodes opening during bending (no dielectric)
: Average gap of the electrodes opening during bending (with dielectric)
: Average gap of the electrodes opening during bending at the dielectric area.

Acknowledgments

The work was supported by Research University Grant, Universiti Sains Malaysia, under Grant 1001/PMEKANIK/814099 and Graduate Research Fund, under Grant 308/AIPS/415401. The authors also express their gratitude for the facilities and opportunities provided by Collaborative Microelectronic Design Excellence Centre USM (CEDEC) in undertaking this project.