Abstract

This research is looking for three laminated composite material groups. These three groups were utilized in experimental investigation to find their mechanical properties. These properties have been used to design and manufacture a socket for a partial foot prosthesis using an ANSYS model. This socket was manufactured with a vacuum pressure device to improve its properties. The socket composite material was tested for tensile and fatigue properties; then, its results were used in the ANSYS model. The composite material matrix was laminated in an 80 : 20 ratio, and there were three types of reinforcement lamination material (Perlon, glass fiber, and carbon fiber). The mechanical property results of these tests were found as follows: using only-Perlon reinforcement, the properties are , , and ; using (3Perlon +2carbon fiber +3perlon) layers, the properties were , , and ; and using (3Perlon + 2 glass fiber + 3perlon) layers, the results were , , and . The ANSYS model used the boundary condition from the measured contact pressure between the socket and the patient’s stump. The MatScan (F-socket) pressure sensor utilized these interface pressure measurements. The maximum values for the pressure were found as follows: 190 kPa and 164 kPa, which are recorded in the posterior and lateral locations, respectively. The calculated factor of safety for the prosthesis that has been made from a selected composite material with the following layers (3 Perlon+2 carbon fiber+3 Perlon) is 1.037 which is safe for design prosthetic applications. From this study, more prosthetic designs can be modelled and manufactured using this approach. Prosthetics and orthotics are usually custom-made for each patient according to its specific requirements. So, it will be very helpful to find a procedure to analyze the prosthetics before manufacturing it.

1. Introduction

Partial foot amputation is a foot amputation between the ankle joint and the lower distal limb. Prostheses for partial foot amputations vary in design and function. Partial foot amputation is a type of amputation, which has many challenges for the rehabilitation physicians, the prosthesis technicians, and the patient himself in comparison to other types of amputations due to the complexity of this type of amputation. The structure of the partial foot prosthesis should be strong to protect the amputation edges from the impact at heel strike and toe off, which should also be able to control the foot deformation in dorsiflexion, plantar flexion, eversion, and inversion. Also, the prosthesis should simulate a suitable gait pattern which can allow the patient to perform normal activities. Then, the prosthesis appearance should be cosmetically acceptable. There are a wide range of partial foot amputation prostheses which can be simple as bunched up socks or foam toe fillers. They also can be a sculptured Plastazote, elastomers2, or orthoprosthesis which can have some extension that reaches to the patella tendon-bearing area [1]. As shown in Figure 1, the patient condition with a foot amputation has different names, such as transmetatarsal amputation, Chopart amputation, Lisfranc amputation, Syme’s amputation, and ankle disarticulation. From these names, the amputation position through the foot can be located; some of these names are from the surgeons who deliver these amputations for the first time such as Chopart, Lisfranc, and Syme. Partial foot amputation (PFA) is defined as losing a part from the foot, either the forefoot or the hindfoot [2], and it is typically due to diabetes [35]. Also, frostbite trauma and congenital anomalies can be another example for other reasons for amputation [6]. During the 19th century, François Chopart was employed at a charitable hospital in Paris, and he defined the disarticulation through the midtarsal joint [7]. At that time, this operation was thought to be having very limited applications due to the fact that the remaining of the foot is going to progress to equinovarus deformity. As time passed, more people are in favour of Chopart amputation because of its ability to keep the limbs’ total length and the high percentage of success during this procedure. An improved high-topped shoe was introduced to assist patients with apropulsive ambulation due to the combination of ankle fusion and hindfoot amputation [811]. Jweeg et al. [12] used different layers of Perlon. Abbas et al. [13] worked on a composite material of carbon fiber with Perlon and Perlon only. Research to investigate KAFO made with a carbon fiber material was performed by Takhakh and Abbas [14].

Abbas investigates the effects of composite material layers on the prosthetic mechanical properties for partial foot amputation [15]. Abbas et al. investigate the mechanical properties of a prosthesis designed for partial foot amputation [16]. Takhakh et al. studied the different types of composite material lamination layers for prosthetic applications [17]. S. M. Abbas and M. H. Abbas used carbon fiber with four layers to make a composite material for an above-the-knee amputation socket [18]. Salman et al. studied a composite material with a wide range of fiber reinforcements [19]. Also, Mostafa et al., Sharba et al., and Salman et al. investigate the hybrid composite material and found its mechanical properties with biaxial tensile, compression, and fatigue tests [2022].

The aim of this work is to design and manufacture a prosthesis for Chopart amputation. This prosthesis is using one of the three composite material groups. The selected composite material is processed by simulating the prosthetic model for these three groups using the mechanical properties collected from the tests that were utilized in this study. The aim of the study was achieved by the following phases.

In the beginning, three types of lamination samples were tested in tensile and fatigue tests, and they are 3 Perlon-2 carbon fiber-3 Perlon, 3 Perlon-2 glass fiber-3 Perlon, and 6 Perlon with a matrix of lamination (80 : 20) resin. Then, a study of the mechanical properties for these materials was done to select the best option among them to make a socket with the required performance. This study was utilized by finite element analysis using ANSYS (14.5) to determine the fatigue safety factor numerally.

The experimental and numerical data used to support the findings of this study are included within the article.

2. Experimental Procedures

The materials for Chopart amputation prostheses used in this study are cotton Perlon, carbon and glass fibers, and resin (80 : 20) with a hardener [23] as shown in Figure 2.

The tools and devices used for this research are listed as follows: (1)A mould from Jepson(2)Pressure vacuum to prevent gaps between the fiber and the resin(3)Tensile test machine(4)Fatigue test for a flat specimen suitable for the composite material

Three samples were tested for each type of composite material in the tensile test device. These samples were prepared according to ASTM D638 type I [24] as shown in Figure 3(a). However, its thickness varies due to the type of lay-up.

Ten samples for each group combination were utilized for the fatigue test. The samples are shown in Figure 3(b); these samples are made according to the standard of the fatigue device while thickness of the samples varies with each group.

The following steps are measuring the partial foot amputation socket dimensions for the patient to manufacture a socket. After manufacturing, the alignment of the partial foot prosthesis was done using the alignment tools to be ready to use by the patient.

The procedure to manufacture a socket can be listed in the following steps: (1)Measuring the following parameters from the patient:(a)The circumferences of the stump(b)Mediolateral diameter at the calcaneus and malleolus(c)The amputee’s length(d)The normal foot’s length(e)The patient shoe’s heel height(2)Handing casting(3)Preparing for rectification and cast rectification(4)Soft socket manufacturing(5)Adding lamination for the socket(6)Trimming and then finishing

Measure the interface pressure using an F-socket at different regions of the patient who wears a prosthetic type PF (age (55 years), height (190 cm), and weight (88 kg)). The patient suffered from left foot amputation (Chopart amputation) due to an explosion. The interface pressure was taken by using a sensor type (MatScan), shown in Figure 4, which can be used for this type of dynamic load applications.

3. Numerical Analysis

In this research, a finite element method was used to evaluate the fatigue characteristics of the prosthetic part. The prosthesis geometry was measured and modelled in ANSYS software as shown in Figure 5. This model used the collected mechanical properties from the tests of the three groups of the composite material, which were used to make the socket. The mechanical properties of the three different groups were used to evaluate the stresses that were generated in the prosthetics due to the weight and walking loading.

The prosthesis displacement boundary conditions and the applied forces are shown in Figure 6.

The software analyzes the prosthesis by creating elements and nodes as shown in Figure 7.

The boundary conditions for this model can be classified into displacement and applied force or pressure. The displacement boundary conditions are in the bottom of the foot where the foot is in contact with the ground. On the other hand, the force and pressure applied to the prosthetic model are from the weight of the patient (88 kg) and from the measured interface pressure from the MatScan sensor. The model shows how the prosthesis will respond to the applied boundary conditions above after solving this model.

4. Results and Discussion

4.1. Tensile Test Results

The specification of the three composite lamination layer groups is in Table 1.

The mechanical properties for these three lamination layer groups are listed in Table 2. The stress-strain curve for each lamination group is shown in Figure 8. These results were collected from the tensile test data for this study.

The found data conclude that the yield strength of group B has increased by 16% when two layers of glass fiber were added to the lamination in group C; also, the ultimate tensile strength has increased by about 24% and by 26%. When comparing the results from groups A and C, it was revealed that using two more layers of carbon fiber can increase by 47%, by 62%, and by 48%. The increase in these mechanical properties in groups A and B is due to the mechanical properties of the glass fibers and the carbon fibers added which are higher than that of Perlon.

4.2. Fatigue Property Results

The fatigue tests were utilized for the three lamination groups in this research. The fatigue tests were done using an alternative bending testing machine with the specimen shown in Figure 3(b). The results from the fatigue testing equipment show the specimens’ fracture testing stress with a total number of cycles when the fracture happened as shown in Figure 9.

Results in Figure 9 show that the amount of cycles in fatigue failure is increasing with decreasing fatigue failure stress. All the tests were conducted at room temperature. Each test was repeated three times, and the results shown in Figure 9 are the average for these data.

The collected data from the tensile and fatigue tests are used in modelling the ANSYS finite element model to compare its results.

4.3. Interface Pressure Result

The contact or interface pressure between the socket and the patient’s stump was measured with a MatScan sensor as shown in Figure 4. The sensor was located on the anterior, posterior, lateral, and medial regions, and their results are shown in Figures 1013. In Table 3, the maximum pressure values of the socket and its position are listed. The higher pressure value was recorded at the posterior area with 190 kPa, and the pressure value at the lateral area is 164 kPa. This pressure distribution in the posterior and lateral muscles is due to the extra activity in these muscles during walking to avoid extra pressure in the area of tibia bone (medial region and anterior region). This type of contact pressure distribution is very common with patients having this type of amputation.

The collected data from the interface contact pressure are used as the boundary condition to solve the ANSYS finite element model.

4.4. Numerical Results

The analysis of the partial foot type Chopart prosthetic model was analyzed by finite element using computer ANSYS software, which can find the total deformation, resultant stress (von Mises stress), and fatigue safety factor.

The data show the factor of safety for the three types of the partial foot prosthetic model, which are presented in Figures 1416, respectively. The von Mises stress results of the prosthetic three groups are shown in Figures 1719. The total deformation is shown in Figures 2022 which are for the three groups of the lamination composites.

The factor of safety for group A is 1.037 which is acceptable for this type of applications.

5. Conclusions

(1)The material selection procedure for the prosthesis used in this research can be used in a wide range of prosthetics and orthotics, as most of these applications are custom-made for each patient due to the special amputation specifications, body dimensions, patient’s weight, and other factors. So, by modelling the proposed prosthesis, it can find what material is the best for the patient’s requirements from the total deflection to the factor of safety from the model results(2)There is good improvement in the mechanical properties of the laminations in groups A and B compared to group C due to adding two layers of glass fiber and carbon fiber, respectively. This improvement can be noticed in both the equivalent stress and the fatigue life of the prosthesis from the finite element model(3)The maximum contact or interface pressure values measured on the prosthesis are as follows: at the lateral area, it was 164 kPa, and at the posterior area, it was 190 kPa. These maximum pressure values are due to the extra activity of the patient’s muscles in this area during walking compared to the anterior regions and medial region which lead to reduced pressure on the tibia area. This contact pressure distribution should be noticed in a wide range of prostheses. Otherwise, high interface contact pressure on the tibia will make the patient uncomfortable and hurt the patient with time(4)In this research, the selected composite material for the prosthetic model shows that the fatigue factor of safety for group A is 1.037 which can be considered safe for this type of application

Data Availability

The experimental and numerical data used to support the findings of this study are included within the article.