Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters
Table 5
Bland-Altman limits of agreement analysis.
| Sodium Comparison | Test | Ref. | Test | Control | Difference (Test - Control) | B-A Limits of Agreement | Equivalent 4 mmol/L Allowable Error | Mean | SD | Mean | SD | Mean | SD | -value | LB | UB |
| A2 | Control | 139.75 | 2.03 | 139.75 | 2.37 | 0.00 | 1.54 | 1.000 | −3.02 | 3.02 | Equivalent | B2 | Control | 139.46 | 2.15 | 139.75 | 2.37 | −0.29 | 1.61 | 0.355 | −3.44 | 2.86 | Equivalent | C2 | Control | 139.61 | 1.93 | 139.75 | 2.37 | −0.14 | 1.51 | 0.620 | −3.10 | 2.81 | Equivalent | D2 | Control | 139.74 | 2.35 | 139.82 | 2.39 | −0.07 | 1.75 | 0.828 | −3.51 | 3.36 | Equivalent |
| Creatinine Comparison | Test | Ref. | Test | Control | Difference (Test - Control) | B-A Limits of Agreement | Equivalent 5% Allowable Error | Mean | SD | Mean | SD | Mean | SD | -value | LB | UB |
| A2 | Control | 0.82 | 0.15 | 0.81 | 0.15 | 0.01 | 0.07 | 0.602 | −0.13 | 0.15 | Equivalent | B2 | Control | 0.81 | 0.17 | 0.81 | 0.15 | 0.00 | 0.08 | 1.000 | −0.15 | 0.15 | Equivalent | C2 | Control | 0.81 | 0.15 | 0.81 | 0.15 | 0.00 | 0.06 | 1.000 | −0.12 | 0.12 | Equivalent | D2 | Control | 0.80 | 0.16 | 0.82 | 0.15 | −0.02 | 0.08 | 0.161 | −0.18 | 0.13 | Equivalent |
|
|