Clinical Study

A Practical Approach for a Wide Range of Liver Iron Quantitation Using a Magnetic Resonance Imaging Technique

Figure 2

Results of curve fit from a patient with short values by dual acquisition method (Figure 1(a)). Data was acquired by the long TE protocol (Figure 1(b)). The same patient data was acquired by the short TE protocol, there were more data points in the curvature area. It is obvious that data acquired by short TE protocol generated better fit for short case.