Research Article

Effect of Different Breast Densities and Average Glandular Dose on Contrast to Noise Ratios in Full-Field Digital Mammography: Simulation and Phantom Study

Table 3

Mean CNR for each mammary gland tissue ratio adjusted by the AGD as a covariate.

Breast phantom model (Fibroglandular/adipose ratio)CNRUnadjusted difference [95%CI]pCNRAdjusted for AGD difference [95%CI]p

A (30/70)15.73-1.30.627416.14-0.920.1459
B (50/50)14.43[-6.84, 4.24]15.22[-2.23, 0.39]
C (70/30)15.73-2.170.421716.14-2.660.0017
A (30/70)13.57[-7.7, 3.37]13.48[-4.03, -1.29]
B (50/50)14.43-0.870.746515.22-1.740.0006
C (70/30)13.57[-6.4, 4.67]13.48[-2.5, -0.98]