Research Article

Dosimetry and Comparison between Different CT Protocols (Low Dose, Ultralow Dose, and Conventional CT) for Lung Nodules’ Detection in a Phantom

Table 2

Estimated equivalent doses.

ProtocolkVmAEquivalent entrance skin dose/uncertainty (mSv)Equivalent dose to the lung/uncertainty (mSv)

Ultralow dose80300.5/0.060.6/0.06
Low dose (lower mA)120201.32/0.261.32/0.12
Low dose (lower kV)100301.34/0.31.5/0.3
Conventional low dose120301.8/0.362.0/0.16
Conventional HRCT120Automatic9.0/0.89.0/0.58