Research Article

A Review of Doses for Dental Imaging in 2010–2020 and Development of a Web Dose Calculator

Table 8

Minimum, mean, and maximum values of effective and organ (bone marrow, brain, salivary gland, and thyroid) doses from cone-beam computed tomography procedures by age group (children and adults) and field-of-view area (FOV width × height, cm2) reported in the selected publications.

FOV area (cm2) age groupEffective dose (μSv)Organ dose (μGy)
Bone marrowBrainSalivary glandThyroid

Small (<50)Children (n = 11)Mean131.3292.4192.02568.5281.4
Min17.115.214.4336.338.6
Max220.21052.0760.04352.0751.0
Adults (n = 9)Mean53.8226.685.81271.1349.3
Min19.021.018.0523.030.0
Max105.0400.0290.02830.01252.0

Medium (50–150)Children (n = 7)Mean144.3512.2115.92289.21089.1
Min47.931.728.61028.0270.6
Max298.01478.0219.04204.02772.0
Adults (n = 25)Mean105.2172.0204.32094.7694.4
Min28.06.92.832.08.4
Max265.0621.01155.06372.03533.0

Large (>150)Children (n = 4)Mean177.7346.62572.92093.91498.6
Min64.7104.7772.21319.0202.6
Max392.2614.04324.04007.02908.0
Adults (n = 20)Mean146.1241.9837.63046.01221.5
Min31.010.0124.0738.347.0
Max368.01034.04653.07775.08727.0