Can Dental Office Lighting Intensity Conditions Influence the Accuracy of Intraoral Scanning?
Table 2
Data obtained when comparing for trueness.
Scan #
Lux
Std. Dev.
Min.
Max.
Avg.
N.C.1
400
0.0419
-0.1042
0.3016
0.0111
N.C.2
400
0.0357
-0.0859
0.2964
0.0133
N.C.3
400
0.0398
-0.1279
0.3204
0.0149
N.C.4
400
0.0393
-0.134
0.2805
0.0113
N.C.5
400
0.0401
-0.1344
0.2592
0.0149
L.N.1
1000
0.0508
-1.4853
0.2996
0.0144
L.N.2
1000
0.0376
-0.1489
0.2378
0.0111
L.N.3
1000
0.038
-0.1193
0.257
0.013
L.N.4
1000
0.0389
-0.1337
0.2262
0.0159
L.N.5
1000
0.0402
-0.1071
0.2792
0.0181
U.J.1
3300
0.0371
-0.1145
0.2422
0.0128
U.J.2
3300
0.0387
-0.0945
0.2474
0.0197
U.J.3
3300
0.0431
-0.135
0.2526
0.0191
U.J.4
3300
0.0377
-0.0925
0.2468
0.0173
U.J.5
3300
0.033
-0.0951
0.2522
0.0117
U.J.N.1
3800
0.0465
-0.1461
0.2364
0.021
U.J.N.2
3800
0.0336
-0.089
0.3087
0.0132
U.J.N.3
3800
0.0337
-0.0987
0.2396
0.0118
U.J.N.4
3800
0.0379
-0.0839
0.2393
0.0161
U.J.N.5
3800
0.0387
-0.1117
0.2435
0.015
U.F.1
10000
0.0317
-0.0924
0.2388
0.012
U.F.2
10000
0.0383
-0.0908
0.2368
0.0141
U.F.3
10000
0.0352
-0.0887
0.2252
0.0169
U.F.4
10000
0.0403
-0.1073
0.2517
0.0166
U.F.5
10000
0.0334
-0.1077
0.2428
0.0094
U.F.N.1
11000
0.0479
-0.1066
0.2286
0.0231
U.F.N.2
11000
0.0303
-0.0792
0.2297
0.0091
U.F.N.3
11000
0.0284
-0.0773
0.2354
0.0092
U.F.N.4
11000
0.0382
-0.0994
0.2459
0.0154
U.F.N.5
11000
0.0392
-1.3395
0.2524
0.0175
Scan #: the name of the STL file and subsequent mesh that was analyzed; Lux: light intensity under which the specific scan was performed measured in lux; Std. Dev.: the standard deviation of all the gap distance values measured in mm; Min.: the smallest gap distance value measured in mm; Max.: the largest gap distance value measured in mm; Avg.: the arithmetic mean of all the gap distances measured in mm.