Can Dental Office Lighting Intensity Conditions Influence the Accuracy of Intraoral Scanning?

Table 2

Data obtained when comparing for trueness.

Scan #LuxStd. Dev.Min.Max.Avg.

N.C.14000.0419-0.10420.30160.0111
N.C.24000.0357-0.08590.29640.0133
N.C.34000.0398-0.12790.32040.0149
N.C.44000.0393-0.1340.28050.0113
N.C.54000.0401-0.13440.25920.0149
L.N.110000.0508-1.48530.29960.0144
L.N.210000.0376-0.14890.23780.0111
L.N.310000.038-0.11930.2570.013
L.N.410000.0389-0.13370.22620.0159
L.N.510000.0402-0.10710.27920.0181
U.J.133000.0371-0.11450.24220.0128
U.J.233000.0387-0.09450.24740.0197
U.J.333000.0431-0.1350.25260.0191
U.J.433000.0377-0.09250.24680.0173
U.J.533000.033-0.09510.25220.0117
U.J.N.138000.0465-0.14610.23640.021
U.J.N.238000.0336-0.0890.30870.0132
U.J.N.338000.0337-0.09870.23960.0118
U.J.N.438000.0379-0.08390.23930.0161
U.J.N.538000.0387-0.11170.24350.015
U.F.1100000.0317-0.09240.23880.012
U.F.2100000.0383-0.09080.23680.0141
U.F.3100000.0352-0.08870.22520.0169
U.F.4100000.0403-0.10730.25170.0166
U.F.5100000.0334-0.10770.24280.0094
U.F.N.1110000.0479-0.10660.22860.0231
U.F.N.2110000.0303-0.07920.22970.0091
U.F.N.3110000.0284-0.07730.23540.0092
U.F.N.4110000.0382-0.09940.24590.0154
U.F.N.5110000.0392-1.33950.25240.0175

Scan #: the name of the STL file and subsequent mesh that was analyzed; Lux: light intensity under which the specific scan was performed measured in lux; Std. Dev.: the standard deviation of all the gap distance values measured in mm; Min.: the smallest gap distance value measured in mm; Max.: the largest gap distance value measured in mm; Avg.: the arithmetic mean of all the gap distances measured in mm.