Research Article

In Vivo Magic Angle Magnetic Resonance Imaging for Cell Tracking in Equine Low-Field MRI

Figure 5

Differences in signal-to-noise ratio (SNR) between SPIO or “false” artefacts. Boxplots exhibiting the comparison of artefact SNR in images categorized as 1 (“hypointense area most likely related to SPIO”=SPIO artefact) and 2 (“questionable hypointense area”=“false” artefact) in T1-weighted sequences obtained at or magic angle (). It shows that the artefact SNR of the ROI (a) and area (b) in the potentially false-positive category 2 images was not as low as in the category 1 images (). However, a threshold SNR value for the discrimination of true SPIO artefacts and other hypointense artefacts was not definable. Circles and rhombs display outlier values; different assessment time points are summarized.
(a)