Tab 5
/
Research Article
Absorbed Radiation Dose in Radiosensitive Organs Using 64- and 320-Row Multidetector Computed Tomography: A Comparative Study
Table 5
The effect of heart rate variation on the absorbed radiation dose (mGy) for both scanners.
Modes
60 bpm
75 bpm
-value
320-row volume mode
15.50
27.59
<0.001
64-row helical mode
22.25
60.24
<0.001