Research Article

Simultaneous Quantification of Dexpanthenol and Resorcinol from Hair Care Formulation Using Liquid Chromatography: Method Development and Validation

Table 4

Results of accuracy study.

Solution% of nominal
value for RC
Concentration of final
solution in µg·mL−1
Recovery
%
Observed concentration of
final solution (µg·mL−1)
Average assay value
(mg·100 mL−1)
Observed
recovery
%
Accuracy
%
% RSD

10040.0040.3350.420.20
RA80 32.0080.0032.2740.3580.03100.03
RB110 44.03110.0744.9356.17111.40101.210.82
RC120 48.12120.3048.3360.42119.8399.61

Solution% of nominal
value for DP
Concentration of final
solution in µg·mL−1
Recovery
%
Observed concentration of
final solution (µg·mL−1)
Average assay value
(mg·100 mL−1)
Observed
recovery
%
Accuracy
%
% RSD

10040.0039.8249.770.02
DA80 32.0080.0031.8839.8479.9199.89
DB110 44.12110.3044.0555.09110.69100.350.48
DC120 47.99119.9747.6959.61119.7799.83

Note: average from three replicate injections from sample preparations.
The recovery study is carried out over three-concentration range.