Abstract

To explore the effect of the full iterative model reconstruction algorithm (IMR) on chest CT image processing and its adoption value in the clinical diagnosis of lung cancer patients, multislice spiral CT (MSCT) scans were performed on 96 patients with pulmonary nodules. Reconstruction was performed by hybrid iterative reconstruction (iDose4) and IMR2 algorithms. Then, the image contrast, spatial resolution, density resolution, image uniformity, and noise of the CT reconstructed image were recorded. The benign and malignant pulmonary nodules of patients were collected and classified into malignant pulmonary nodule group and benign pulmonary nodule group, and the differences in chest CT imaging characteristics between the two groups were compared. The subject’s receiver operating characteristic (ROC) curve was used to analyze the diagnostic sensitivity, specificity, and area under the curve (AUC) of CT for benign and malignant pulmonary nodules. It was found that the spatial resolution, density resolution, image uniformity, and contrast of the CT image reconstructed by the IMR2 algorithm were remarkably greater than those of the iDose4 algorithm, and the noise was considerably less than that of the iDose4 algorithm (). Among 96 patients with pulmonary nodules, 65 were malignant nodules, including 15 squamous cell carcinoma, 31 adenocarcinoma, and 19 small cell carcinomas. There were 31 cases of benign nodules, including 14 cases of hamartoma, 10 cases of tuberculous granuloma, 2 cases of sclerosing hemangioma, and 5 cases of diffuse lymphocyte proliferation. The pulmonary nodule malignant group and the pulmonary nodule benign group had statistical differences in pulmonary nodule size, nodule morphology, burr sign, lobular sign, vascular sign, bronchial sign, and pleural depression sign (). The sensitivity, specificity, and area under the curve (AUC) of IMR2 algorithm processing chest CT images for liver cancer diagnosis were 85.7%, 82.3%, and 0.815, respectively, which were significantly higher than the original CT images (). In short, chest MSCT based on the IMR2 algorithm can greatly improve the diagnosis efficiency of lung cancer and had practical significance for the timely detection of early lung cancer.

1. Introduction

Lung cancer is a kind of malignant tumor that occurs in the lungs. It is one of the most common malignant tumors in the world, and its incidence ranks first in the cause of death among tumor diseases in China. Lung cancer is classified into non-small cell lung cancer (NSCLC) and small cell lung cancer (SCLC) according to the pathological type. Among them, NSCLC includes adenocarcinoma (AC), squamous cell carcinoma (SCC), and large cell carcinoma (LCC). NSCLC accounts for more than 85% of all lung cancers [1, 2]. At present, lung cancer is considered as a respiratory disease, and about 80% of lung cancers are mainly caused by smoking. Global air pollution is also one of the main causes of lung cancer-related deaths [3]. The key to reducing lung cancer mortality lies in early detection. However, early lung cancer often has no obvious clinical symptoms, and health examinations are needed to achieve early diagnosis. Therefore, it is necessary to diagnose and treat lung cancer as early as possible.

As one of the main methods for early diagnosis of lung cancer, chest computerized tomography (CT) has the advantages of high spatial resolution, high density resolution, and good anisotropy, but there are also a series of problems that need to be optimized, such as noise and artifacts [4]. In recent years, many advances have been made in the optimization technology for CT images. The reconstruction algorithm for CT images was originally developed by the filtered back projection (FBP) algorithm. FBP algorithm uses filtering function or convolution algorithm to balance the spatial resolution and noise of the final reconstructed image. Due to the high requirements for the amount of projection data, this algorithm makes the CT scan dose reduced, the reconstruction time is too long, and the CT image has problems such as noise [5]. To break through the limitations of the FBP algorithm, people began to try new optimization algorithms, such as adaptive statistical iterative reconstruction (ASiR) [6], iterative reconstruction in image space (IRIS) [7], adaptive iterative dose reduction (AIDR) [8], and hybrid iterative reconstruction technology (iDose).

As an iterative reconstruction algorithm using dual models, iDose4 is part of the iterative technology based on FBP. This algorithm can significantly reduce the noise of CT images, but it is still a part of iterative technology. When image reconstruction is performed, the influence of the system hardware and the photon characteristics of part of the incident X-ray on the image are not considered, which limits the further optimization of CT images [9]. The knowledge based IMR is a new upgrade based on iDose4, which removes the FEP component. The optimization effect is realized through continuously optimizing image statistics, data statistics, and system models in the image and data space. In recent years, a comparative study on the IMR algorithm and the FEP algorithm pointed out that, compared to FEP and IMR, it can reduce the radiation dose by 60% to 80%, reduces noise by 70% to 90%, and increases the spatial resolution by about 50% [10]. It is verified that the IMR algorithm has a good application prospect in the optimization of CT images. At present, the IMR algorithm has been applied to the clinical diagnosis of many diseases, but there is no report about the adoption value of IMR in the early diagnosis of lung cancer patients.

Therefore, the IMR algorithm based on the iDose4 algorithm was employed to iteratively reconstruct the chest CT of lung cancer patients. The receiver operating characteristic (ROC) curve was used to analyze the early diagnosis value of the two algorithms for lung cancer patients, so as to comprehensively evaluate the application value of the IMR algorithm in the CT diagnosis of lung cancer patients. It was hoped to provide some reference for the related research on the optimization of chest CT image quality and the improvement of diagnostic accuracy in clinical lung cancer patients.

2. Materials and Methods

2.1. Research Object

In this study, 96 patients with lung cancer admitted to hospital were selected as the research object, including 54 males and 42 females, aged 36–72 years. All patients received multislice spiral CT (MSCT) scans. The average age of patients was 58.05 ± 10.95 years, the body mass index (BMI) ranged from 19.70–29.64 kg/m2, and the average BMI was 26.35 ± 3.67 kg/m2. The benign and malignant pulmonary nodules of patients were collected and classified into malignant pulmonary nodule group and benign pulmonary nodule group. The study had been approved by the ethics committee of the hospital. The patients included in the study and their family members were informed and signed the informed consent forms.

Inclusion criteria: (i) the chest CT showed a lung mass with a diameter less than or equal to 3 cm, with clear or fuzzy edges, and a circular-like density with increased density; (ii) complete basic clinical information; (iii) patients without serious heart, liver, or kidney dysfunction and serious infection; (iv) patients without previous history of tumors or other types of tumors. Exclusion criteria: (i) patients who had received chemotherapy and radiotherapy, immunotherapy, and other antitumor treatments before surgery; (ii) patients with other malignant tumors; (iii) female patients during menstruation, pregnancy, and lactation.

2.2. CT Scans

MSCT scans were performed on 96 patients with pulmonary nodules. The CT model was a microplate 3D iCT (Philips Corporation, Cleveland, USA). Breath-hold training was made before plain CT scan of the chest. During the scan, the subject was in a supine position with his arms raised and his head advanced. Scanning conditions: 120 Kv, 30 mAs for malignant pulmonary nodule group; 120 Kv, 10 mAs for benign pulmonary nodule group. Collimation was 128 × 0.625 mm, pitch was 0.992, rotation time was 0.75 s/r, and matrix was 512 × 512. The scan range was from the tip of the lung to the bottom of the lung. Scanning parameters: pitch was 0.920, stereoscopic field of view (SFOV) was 50 cm, rotation time was 1 s/r, and collimation was 64 × 0.625 mm. The tube voltage was 120 kV, the tube current was set in 9 groups, and the current sizes were 20, 40, 60, 80, 100, 200, 300, 400, and 500 mAs, respectively.

2.3. Patient CT Optimization Based on iDose4 Algorithm and IMR Algorithm

The original data of the CT images obtained from the scan were reconstructed by the iDose4 algorithm and the IMR algorithm, respectively. The data parameters of iDose4 were Standard (B). iDose was employed to iteratively reconstruct the 4th level among the 7 levels, which was recorded as iDose4. IMR iteratively reconstructed the second level of the 3 levels, which was denoted as IMR 2. The reconstruction parameters were Routine, Soft, and SharpPlus. Display field of view (DFOV) was 25 cm, layer thickness was 5 mm, layer spacing was 5 mm, and matrix was 512 × 512. The reconstructed image data was saved in a disc. Image detection was performed by Iris image quality automatic detection software (V2.4, 2009, Iris QA, LLC, USA).

The IMR algorithm model is upgraded and optimized based on the image degradation model. First, LR observation images are set.

In the model of (1), is an (N = m n) vector composed of k LLR observation images sorted by dictionary. If and are downsampling factors in the horizontal and vertical directions, respectively, then Z is the vector of the HH image formed after dictionary sorting, and is the affine transformation matrix. Bk is the fuzzy matrix, D is the downsampling matrix of , and is the Gaussian white noise of vector, which can be obtained from model (1). The image SSR reconstruction needs to solve the following minimization problems, and the corresponding equation is as follows:

Another balance item is introduced for adjustment, and there is

In (3), is the balance factor and is the regularization term. There are many ways to determine .where and are the movement matrixes that move the image z in the horizontal and vertical directions by l and m pixels, respectively, and (0 <  < 1) is the weighting coefficient.

In short, a constant will have a double effect on the reconstructed HR image, so how to choose a suitable is very important for the quality of the final reconstructed HR image. An adaptive weighting coefficient is introduced. According to the difference between the specific edge and the smooth part, the edge of the image is sharpened, and the other parts of the image are also smoothed and noise is suppressed, so as to effectively ensure the image quality. The specific method is as follows.

It is assumed that (5) can be obtained from (4): shall have a physical meaning. The element in the vector is the difference between the pixel z at the same position in the reconstructed image and the surrounding pixels. In a grayscale image, if it is too large, it means that the reconstructed image z has a sudden change around the i-th pixel; that is, there is an edge around the i point. Therefore, it is hoped that the reconstruction algorithm can sharpen the edge around the i point and highlight the display effect of the edge. Otherwise, it is too small, indicating that the i pixel in the image is very smooth, and it is hoped that the reconstruction algorithm can suppress noise. In summary, to perform SSR reconstruction according to the characteristics of different images, an adaptive bilateral full-variable regularization term is introduced, which is expressed as follows.

shall have a physical meaning. The element in the vector is the difference between the pixel at the same position in the reconstructed image z and the pixels around it. In a grayscale image, , if is too large, it indicates that the reconstructed image z has a sudden change around the first pixel; that is, there is an edge around the point. Therefore, it is hoped that the reconstruction algorithm can sharpen the edge around the ip point to highlight the display effect of the edge. Otherwise, is too small, which means that the i pixel of the image is very smooth, and it is hoped that the reconstruction algorithm can suppress noise. In summary, to realize SSR reconstruction based on the characteristics of different images, an adaptive bilateral full variation regularization term is introduced as follows.

Equation (6) is the adaptive weighting coefficient that changes according to the difference of each pixel of the reconstructed HR image, which fully guarantees the adaptability of the algorithm, thereby theoretically improving the quality of the reconstructed HR image.

The computer automatically analyzed and detected the CT scan data of 96 patients, and the computer gave a test report. Measurement indicators included spatial resolution, density resolution, image uniformity, and noise conditions.

2.3.1. Spatial Resolution Measurement

It was known that the CT of the patient contained 21-line pairs for visual inspection, which were used as the point source for the software to calculate the modulation transfer function (MTF). First, some images were randomly selected during the preliminary experiment, and the difference between the visual inspection method and the MTF method in evaluating the spatial resolution was compared. Giving that the corresponding value of MTF10% was close to the visual method, it was set as the reference value of image spatial resolution. In this study, if the tube current was less than 50 mAs, since the software cannot automatically detect the point source, it would be sent to the postprocessing workstation. The data cannot be collected accurately, so visual inspection was used to record the line logarithm. This operation was jointly completed by two radiologists with more than 5 years of work experience, and the differences in the judgment results were negotiated and handled.

2.3.2. Density Resolution Measurement

In the Iris image quality automatic detection report, the density resolution is expressed by contrast, which is the minimum CT value required to distinguish a target of a certain size. In this experiment, the contrast of the CT value required to distinguish a target with a diameter of 5 mm was included as an objective indicator of the density resolution (unit: %).

2.3.3. Image Uniformity and Noise Measurement

In the chest CT image, the software automatically identified and delineated the region of interest and measured the CT value and noise (SD) of the center, upper, and lower left and 2 cm right away from the image edge. The maximum deviation between edge and center CT value was the reference value of image uniformity. The ambient and central noise values were averaged as the noise of this image.

2.4. Acquisition of CT Imaging Omics Features and Detection of Optimized Effects in Patients with Lung Cancer

Imaging analysis was performed on the optimized CT images of lung cancer patients. The statistics on the typical imaging characteristics of lung cancer patients were collected, and the malignant group of pulmonary nodules and the benign group of pulmonary nodules were analyzed in terms of the imaging characteristics of the CT images of the two groups of patients. Moreover, the differences in the imaging characteristics of the CT images of the two groups of patients were compared, including (i) nodule size: the maximum axial diameter of the nodule in the mediastinal window was measured; (ii) nodule morphology: regular or irregular; (iii) fuzzy sign: it was manifested as a long line shadow or short dense thin line shadow that extended radially from the edge of the nodule to the surrounding lung field in the lung window and was not connected to the pleura; (iv) lobular sign: manifested as obvious unevenness on the edge of the nodule, and multiple arcs with notches; (v) vascular sign: it was manifested as the vascular structure around the nodule gathering in the direction of the lesion, or it was manifested as the blood vessel passing through or ending at the lesion; (vi) bronchial sign: mainly manifested as thin strips, tubular shadows with gas-like density, or dot-like translucent shadows appearing on continuous layers; (vii) cavitation sign: mostly 2-3 mm gas sample density shadows, excluding the density reduction shadows that appeared on continuous layers, which may be unoccluded small bronchi or alveoli that became larger lumens or vacuoles due to traction; (viii) pleural depression sign: manifested as a linear depression or triangular depression between the tumor lesion and the pleura.

Then, the ROC curve was used to analyze the differences in the sensitivity, specificity, and AUC area of the two algorithms in the diagnosis of benign and malignant pulmonary nodules, so as to measure the optimization value of the algorithm.

Then, the sensitivity, specificity, and AUC of the two algorithms in the diagnosis of benign and malignant pulmonary nodules were analyzed by ROC curve, so as to measure the optimization value of the algorithm. The sensitivity and specificity are calculated by the following equations, respectively:

A is the degree of coincidence between fitting the real lung cancer lesion area (set as A) and the lung cancer lesion area determined by the full model iterative algorithm (set as C). B and D represent other positions outside the real cancer lesion area of lung cancer and other positions outside the cervical cancer lesion area determined by artificial intelligence algorithm, respectively. In addition, three overlapping tumor regions were divided into different structural groups of lung cancer tumors that were, whole-area tumors (WT), core tumor areas (CT), and enhanced tumor areas (ET).

2.5. Statistical Methods

SPSS 19.0 was employed for data statistics and analysis. Percentage (%) was how count data were expressed, and mean plus or minus standard deviation was how measurement data were expressed. The comparison of the means between the groups was performed by t-test. The count data was expressed as a percentage (%), and the test was used. The Mann–Whitney U test was used to compare the differences in the macroscopic characteristics and imaging omics characteristics of the two types of lung cancer patients, and indicated that the differences were statistically significant.

3. Results

3.1. Comparison of CT Images Processed by Different Algorithms

Figure 1 was a chest CT image of a 67-year-old patient. The noise of CT images of lung cancer patients after iDose4 algorithm processing was improved in contrast to the CT images of patients before processing, and the image clarity was also improved to a certain extent. The CT image processed by the IMR 2 algorithm was significantly improved compared to the one before the processing. In addition, the IMR algorithm had a better processing effect.

3.2. Evaluation of CT Image Quality Based on Different Algorithms

Figures 25 show the comparisons of the spatial resolution, density resolution, image uniformity, and noise of the iDose4 algorithm and the IMR algorithm under different parameters, respectively. The spatial resolution, density resolution, image uniformity, and contrast of the CT image reconstructed by the IMR2 algorithm were significantly greater than those of the iDose4 algorithm, and the noise was notably less than that of the iDose4 algorithm (). Through analysis of variance (ANOVA) test, iDose iterative reconstruction cannot improve the spatial resolution of CT images when the reconstruction parameter was Standard (B). The spatial resolution detection results of IMR and iDose were dramatically different (). The spatial resolution of IMR 2 was improved by an average of 16% in contrast to that of iDose4. In IMR iterative reconstruction, the spatial resolution of SharpPlus was greatly higher than that of Routine and Soft, and the spatial resolution of Routine was remarkably superior to that of Soft. The differences among the three groups were considerable through ANOVA test ().

3.3. Comparison of CT Image Optimization Effects of Two Algorithms

Based on the detected sensitivity, specificity, and AUC, the optimization effect of the patient’s CT image after the iDose4 algorithm and IMR2 algorithm reconstruction and optimization processing is shown in Figures 6 and 7. The sensitivity, specificity, and AUC of the patient’s diagnosis of benign and malignant pulmonary nodules before CT image processing were 59.4%, 68.1%, and 0.632, respectively. The diagnostic sensitivity, specificity, and AUC of CT images reconstructed by iDose4 algorithm for benign and malignant lung nodules were 82.1%, 75.3%, and 0.695, respectively. Those of CT images reconstructed by IMR2 algorithm for benign and malignant pulmonary nodules were 85.7%, 82.3%, and 0.815, respectively, which were significantly higher than the original CT images ().

3.4. Comparison of CT Imaging Information of Two Groups of Patients Processed by IMR 2 Algorithm

The CT images and other information of 96 patients with lung cancer are summarized in Figures 810. Figure 8 shows the distribution of lung cancer types, Figure 9 shows the comparison of tumor size between the two groups, and Figure 10 shows the comparison of CT imaging features between the two groups. There were 65 cases of malignant nodules in lung cancer patients, including 15 cases of squamous cell carcinoma, 31 cases of adenocarcinoma, and 19 cases of small cell carcinoma. Of which, 31 cases were benign nodules, including 14 hamartomas, 10 tuberculous granulomas, 2 sclerosing hemangiomas, and 5 diffuse lymphocyte hyperplasia. The average size of lung nodules in the malignant group and benign group of lung cancer patients was 23 (15, 27) mm and 14 (11, 19) mm, respectively. The size of benign nodules was significantly smaller than that of malignant nodules (). In the malignant group of lung nodules, the morphology of the nodules was irregular. The proportions of burr sign, lobular sign, vascular sign, bronchial sign, and pleural depression sign were 92%, 48%, 42%, 45%, 38%, and 43%, respectively. In addition, the percentages of indications in the benign pulmonary nodule group were 59%, 29%, 19%, 27%, 18%, and 20%, respectively. There were significant differences in the number of patients with each indication between the two groups () (Figure 10).

4. Discussion

Chest CT lung cancer screening is the best way to detect early lung cancer [11]. At present, the use of chest CT for preoperative diagnosis of lung cancer is relatively common, but due to the large image noise and other shortcomings, the traditional method of using the subjective judgment of physicians is likely to be misdiagnosed [12, 13]. Therefore, how to accurately locate the tumor lesions still needs further research. The advent of iterative reconstruction algorithms has greatly reduced image noise. With the advancement of computer software and hardware technology, it has been widely used clinically. In the evaluation system of CT image quality, image noise (SD, SNR, and CNR) is only one aspect, and its core evaluation indicators are the spatial resolution and density resolution of the image. Studies found that the image quality of images reconstructed by iDose reconstruction and conventional dose scan FBP reconstruction has no difference in image quality and lower noise [14]. In addition, there were reports that the IMR algorithm [15], as an emerging full model iterative reconstruction technology, was mainly used in computer-aided medical imaging diagnosis and analysis to evaluate the efficacy of radiotherapy, image-based pathological analysis, tumor microenvironment, and heterogeneity research. For example, the high-throughput processing of multimodal images and the analysis of graphics into data can provide quantification of diagnosis basis and have powerful image noise reduction capabilities [16, 17]. Based on traditional chest CT, the image quality using the IMR algorithm is good and the noise is low. Therefore, LDCT scan (120 kV, 30 mAs) was adopted in this research, and the image reconstructed by iDose4 was set as a reference to compare and study the ability of IMR iterative reconstruction algorithm to improve image quality. Moreover, the chest CT accuracy, spatial resolution, density resolution, image uniformity [18], and noise were evaluated.

The results showed that the spatial resolution, density resolution, image uniformity, and contrast of the CT image reconstructed by the IMR2 algorithm were significantly greater than that of the iDose4 algorithm, and the noise was significantly less than that of the iDose4 algorithm (). The subject ROC curve was used to analyze the diagnostic sensitivity, specificity, and AUC of benign and malignant pulmonary nodules based on CT. It was found that the diagnostic sensitivity, specificity, and AUC of CT images reconstructed by IMR2 algorithm for benign and malignant lung nodules were significantly higher than that of original CT images (). In short, chest CT based on the IMR2 algorithm can significantly improve the diagnosis of lung cancer and was of practical significance for the timely detection of early lung cancer. This was consistent with the findings of Lombard et al. [19]. In addition, the patients were grouped according to the benign and malignant conditions of different lung nodules, and the differences in chest CT imaging characteristics between the two groups were compared. It was found that the malignant pulmonary nodule group and the benign pulmonary nodule group had statistical differences in lung nodule size, nodule morphology, burr sign, lobular sign, vascular sign, bronchial sign, and pleural depression sign (). It was suggested that the benign and malignant conditions of lung nodules in patients with lung cancer were closely related to the imaging features presented by CT [20]. The tumor size in chest CT images of patients with malignant pulmonary nodules was significantly larger than that in the benign nodules group. The proportion of patients with irregular nodules, burr sign, lobular sign, vascular sign, bronchial sign, and pleural sign was significantly higher than that of the benign group. In summary, chest CT based on the full model iterative reconstruction algorithm had a good diagnostic value for lung cancer patients, can significantly improve the diagnostic efficiency of lung cancer, and was worthy of clinical promotion.

5. Conclusion

To explore the application of IMR processed chest CT images in the clinical diagnosis of lung cancer patients, the iDose4 algorithm was employed and applied to the CT image analysis of 96 lung cancer patients. It was found that the IMR algorithm showed high accuracy and AUC area for lung cancer diagnosis and prediction compared with the iDose4 algorithm for processing patient CT images. Among them, IMR2 algorithm combined with SharpPlus parameters can be used as the optimal feature combination model for the diagnosis of lung cancer. However, the selection of patient samples is small and the source is single in this study, and the different manifestations of lung cancer patients are not discussed in detail. It is impossible to verify the impact of these features on the accuracy of the prediction. In the future, it will consider increasing the sample size of lung cancer patients and further adopting a multicenter collaborative analysis method for research. In conclusion, the results provide good clinical data support and certain theoretical support for the application of the IMR algorithm in the clinical diagnosis of lung cancer.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare no conflicts of interest.

Authors’ Contributions

Jie Li and Wei Wang contributed equally to this work.

Acknowledgments

This work was supported by Zhuhai Science and Technology Plan Medical Project (20181117A010054).