Research Article
An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations
Table 1
The coincidence detection efficiencies () of the different X-rays and gamma rays of xenon radioisotopes by Geant4 modeling.
| Radioxenon | ROI limits, keV | | , % | | -axis (CsI) | -axis (BC404) | ROI index | |
| 133Xe | 17.47–42.59 | 0–347.50 | 1 | 89.6 (87.2) | | 67.80–98.73 | 0–396.50 | 2 | 72.5 (68.5) | | 17.47–42.59 | 16.25–89.55 | 3 | 33.0 (32.6) | | 17.47–42.59 | 239.03–396.504 | 4 | 8.8 (8.0) | | 17.47–42.59 | 170.07–396.505 | 5 | 24.6 (22.9) | | 17.47–42.59 | 16.25–161.90 | 6 | 60.2 (58.4) |
131mXe | 17.47–42.59 | 96.99–161.90 | 7 | 78.0 (74.9) |
133mXe | 17.47–42.59 | 170.07–239.03 | 8 | 69.6 (65.3) | 135Xe | 226.58–275.47 | 0–908.0 | 9 | 72.9 (63.8) |
|
|
Data in brackets is simulated with original PW5.
|