Research Article

An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations

Table 1

The coincidence detection efficiencies ( 𝜀 𝛽 𝛾 ) of the different X-rays and gamma rays of xenon radioisotopes by Geant4 modeling.

RadioxenonROI limits, keV 𝜀 𝛽 𝛾 , %
𝑦 -axis (CsI) 𝑥 -axis (BC404)ROI index

133Xe17.47–42.590–347.50189.6 (87.2)
67.80–98.730–396.50272.5 (68.5)
17.47–42.5916.25–89.55333.0 (32.6)
17.47–42.59239.03–396.50448.8 (8.0)
17.47–42.59170.07–396.505524.6 (22.9)
17.47–42.5916.25–161.90660.2 (58.4)
131mXe17.47–42.5996.99–161.90778.0 (74.9)
133mXe17.47–42.59170.07–239.03869.6 (65.3)
135Xe226.58–275.470–908.0972.9 (63.8)

Data in brackets is simulated with original PW5.