Research Article

The Study of Dosimetric Characteristics of the XHA600D Medical Linear Accelerator Based on a Monte Carlo Code

Table 3

The simulated and measured beam qualities of the FF beam and FFF beam are shown. The uncertainties of the simulated beam qualities of FF and FFF beam were 0.6% and 0.7% (coverage factor kā€‰=ā€‰2), respectively.

Beam
SimulatedMeasuredDiff. (%)

FF0.6810.6770.590
FFF0.6530.6520.153