The Study of Dosimetric Characteristics of the XHA600D Medical Linear Accelerator Based on a Monte Carlo Code

Table 4

Gamma passing rates of off-axis dose curves for different field sizes (cm2) at 10 cm depth in FF mode and FFF mode when the gamma criteria is 1 mm/1% (local dose).

Field size (cm2)FF (%)FFF (%)

4 × 4100100
6 × 6100100
8 × 8100100
10 × 10100100
15 × 15100100
20 × 2099.3100
25 × 2599.697.0
30 × 3099.497.3