The Study of Dosimetric Characteristics of the XHA600D Medical Linear Accelerator Based on a Monte Carlo Code

Table 5

The simulated and measured profiles of penumbras (cm) of FF and FFF beams for different fields at 10 cm depth (The uncertainties of simulated penumbras in FF mode of 4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 cm2 fields were as follows: 0.4%, 0.8%, 0.6%, 0.8%, 1.2%, 1.6%, 1.8%, 2% (k = 2). The uncertainties of simulated penumbras in FFF mode of 4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 cm2 fields were as follows: 0.8%, 0.5%, 0.6%, 0.8%, 1%, 1.2%, 1.4%, 1.6% (k = 2)).

Field size (cm2)FFFFF
Simulated (cm)Measured (cm)Diff. (%)Simulated (cm)Measured (cm)Diff. (%)

4 × 40.6460.649−0.4620.6390.645−0.930
6 × 60.6760.6730.4460.6570.672−2.23
8 × 80.7110.716−0.6980.7020.711−1.27
10 × 100.7480.752−0.5320.7410.752−1.46
15 × 150.8120.8080.4950.7950.804−1.12
20 × 200.8530.867−1.610.8450.866−2.42
25 × 250.8740.913−4.270.8730.902−3.22
30 × 300.9900.991−0.1000.9400.958−1.88