The Study of Dosimetric Characteristics of the XHA600D Medical Linear Accelerator Based on a Monte Carlo Code

Table 6

The simulated and measured values for various field sizes at the depth of for FF and FFF beams. (The uncertainties of simulated values in FF mode of 4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 cm2 fields were as follows: 0.3%, 0.3%, 0.4%, 0.4%, 0.5%, 0.7%, 0.8%, 1% (k = 2). The uncertainties of simulated values in FFF mode of 4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 cm2 fields were as follows: 0.3%, 0.3%, 0.4%, 0.4%, 0.5%, 0.6%, 0.6%, 0.8% (k = 2).).

Field size (cm2)FFFFF
SimulatedMeasuredDiff. (%)SimulatedMeasuredDiff. (%)

4 × 40.9320.933−0.1070.9510.953−0.210
6 × 60.9630.9620.1040.9770.9760.102
8 × 80.9770.984−0.7110.9870.9870.00
10 × 101.001.001.001.00
15 × 151.021.03−0.9711.031.020.980
20 × 201.051.040.9621.041.040.00
25 × 251.061.06−0.001.061.050.952
30 × 301.071.060.9431.071.060.943