The Study of Dosimetric Characteristics of the XHA600D Medical Linear Accelerator Based on a Monte Carlo Code

Table 7

The measured and calculated data of flatness and symmetry of FF beams. (The uncertainties of simulated flatness in FF mode of 4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 cm2 fields were as follows: 0.4%, 0.6%, 0.7%, 0.9%, 1.2%, 1.4%, 1.6%, 2% (k = 2). The uncertainties of simulated symmetry in FF mode of 4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 cm2 fields were as follows: 0.3%, 0.4%, 0.5%, 0.6%, 0.8%, 0.9%, 1%, 1.2% (k = 2).).

Field size (cm2)Flatness (%)Symmetry (%)
Simulated (%)Measured (%)Simulated (%)Measured (%)

4 × 41.772.660.04210.611
6 × 61.822.260.04530.166
8 × 81.571.800.1600.143
10 × 101.491.660.1000.127
15 × 152.021.950.1100.00298
20 × 202.161.470.1300.0303
25 × 252.251.660.3040.598
30 × 302.331.570.2080.626