Intra- and Inter-Rater Reliability of Processing Ultrasound Tissue Characterization Scans in Midportion Achilles Tendinopathy

Table 3

Intra-rater and inter-rater reliability for the processing of UTC scans in quantifying the structure of the Achilles tendon midportion (n =50).

Echo-typeIntra-rater reliabilityInter-rater reliability
ICC (95% CI)SEMMDCICC (95% CI)SEMMDC

Echo-type I%0.98 (0.97–0.99)1.74.70.99 (0.95–0.99)1.23.3
Echo-type II%0.97 (0.94–0.99)0.92.50.98 (0.96–0.99)0.72.1
Echo-type III%0.98 (0.96–0.99)1.33.60.99 (0.95–0.99)0.92.6
Echo-type IV%0.99 (0.97–0.99)0.71.90.99 (0.99–1.00)0.71.9
Echo-type I + II%0.98 (0.96–0.99)2.15.90.99 (0.97–1.00)1.54.2
Echo-type III + IV%0.99 (0.97–0.99)1.54.20.98 (0.95–0.99)2.26.0

Intra-rater and inter-rater reliability are calculated for echo-types I, II, III, and IV individually, and combined for aligned fibrillar structure (echo-types I + II) and disorganized tendon structure (echo-types III + IV). CI: confidence interval; ICC: intraclass correlation coefficient; MDC: minimal detectable change (%); SEM: standard error of measurement (%).