Intra- and Inter-Rater Reliability of Processing Ultrasound Tissue Characterization Scans in Midportion Achilles Tendinopathy

Table 4

Intra-rater and inter-rater reliability for the processing of UTC scans in quantifying the AoMD (n =50).

Echo-typeIntra-rater reliabilityInter-rater reliability
ICC (95% CI)SEMMDCICC (95% CI)SEMMDC

Echo-type I%0.98 (0.97–0.99)1.74.80.97 (0.92–0.98)2.15.9
Echo-type II%0.85 (0.75–0.91)2.98.20.92 (0.86–0.95)2.15.9
Echo-type III%0.96 (0.94–0.98)2.46.80.96 (0.92–0.98)2.46.8
Echo-type IV%0.94 (0.90–0.97)1.74.60.97 (0.95–0.99)1.23.3
Echo-type I + II%0.96 (0.93–0.98)3.59.80.98 (0.95–0.99)2.56.9
Echo-type III + IV%0.96 (0.93–0.98)3.610.00.97 (0.94–0.98)3.18.6

Intra-rater and inter-rater reliability are calculated for echo-types I, II, III, and IV individually, and combined for aligned fibrillar structure (echo-types I + II) and disorganized tendon structure (echo-types III + IV). CI: confidence interval: ICC: intraclass correlation coefficient: MDC: minimal detectable change (%); SEM: standard error of measurement (%); AoMD, area of maximum degeneration (1 slide in the Achilles tendon midportion with the lowest representation of echo-type I).