Table 1

Sampling and analytical conditions.

TD conditionsGC conditionsMS conditions

Configuration: TD-100 for automated tube desorption (Markes International Ltd)
Flow path temperature: 200°C
Focusing trap: air toxics analyser type trap
Trapping temperature: 25°C
Cold trap desorption: 320°C for 10 min (20 : 1 split)
Trap heating rate: 40°C/s
Column: 60 m × 0.32 mm × 1.0 μm, 100% dimethylpolysiloxane
Oven: 35°C (25 min), 5°C/min to 180°C, then 15°C/min to 320°C (12 min)
Constant pressure: 10 psi
Source temperature: 230°C
Quadrupole temperature: 150°C
Transfer line temperature: 280°C
Scan range: m/z 10–400

Sampling: 1 μL of a 100 ng/μL standard loaded onto a multi-bed sorbent tube (Tenax TA, Carbograph 1TD, Carboxen 1003).