Table 8

Sampling and analytical conditions.

TD conditionsGC conditionsMS conditions

Configuration: TD-100 for automated tube desorption (Markes International Ltd)
Flow path temperature: 150°C
Cold trap: air toxics analyser type trap
Trapping temperature: 25°C
Cold trap desorption: 300°C for 3 min (10 : 1 split)
Trap heating rate: 40°C/s
Column: 60 m × 0.32 mm × 1.8 μm, 624 type phase
Oven: 40°C (5 min), 5°C/min to 230°C (0 min)
Constant pressure: 10 psi
Source temperature: 230°C
Quadrupole temperature: 150°C
Transfer line temperature: 280°C
Scan range: m/z 33–350

Sampling: Sample 1 (black trace): 1 L of sample pumped onto multi-bed tubes using a constant-flow pump at 100 mL/min for 10 min; Sample 2 (blue trace): 1 L sampled onto a multi-bed tube, using ten pulls of a piston pump set at 100 mL per pull.