Research Article

Simple Reversed-Phase HPLC Method with Spectrophotometric Detection for Measuring Acetaminophen-Protein Adducts in Rat Liver Samples

Table 2

Results of the intraday variability study on the proposed HPLC method based on three different concentrations of APAP-Cys added to distilled water and to blank rat liver matrix protein dispersiona,b.

APAP-Cys injected (μg)Peak area ratio
(mean ± SD, 𝑛 = 3 )
RSD (%)

Samples in distilled water

0.1560.005 ± 0.0020.91
0.6250.021 ± 0.0010.72
1.2500.042 ± 0.0010.52

Samples in rat liver matrix protein dispersion

0.1560.004 ± 0.0011.02
0.6250.018 ± 0.0010.68
1.2500.040 ± 0.0030.63

aEach sample was analyzed at 4 hr intervals three times on the same day.
bRSD is the relative standard deviation.