Table 6

Parameters of calibrations dependences for the determination of 5-NQ, 2-NF, 1-NP, and 3-NFt using HPLC-ED with “wall-jet” m-AgSAE detector (A) under optimized conditions in deionized water as an ideal matrix and (B) for determination in drinking water after SPE. Mobile phase 0.01 mol L−1 phosphate buffer pH 7.0: methanol (15 : 85; v/v), 𝐹 𝑚 = 1  mL min−1, 𝑉 i n j = 2 0 μL, detection potential −1.5 V, responses evaluated from the peak heights.

AnalyteLinear dynamic range (μmol L−1)SlopeaInterceptbCorrelation coefficient 𝐿 𝐷 (μmol L−1)

(A) Optimized conditions in deionized water

Electrochemical detection5-NQ10–75–1.28–0.6–0.99728.9
2-NF25–100–1.45–2.1–0.97987.8
1-NP25–100–0.91–15.5–0.993512.5
3-NFt25–100–1.001.0–0.998511.4

UV detection5-NQ1–1000.502–0.120.99991.6
2-NF1–1000.423–0.190.99981.9
1-NP1–1000.882–0.390.99990.9
3-NFt1–1000.943–0.120.99990.8

(B) Determination in drinking water after SPE

Electrochemical detection5-NQ0.5–5–25.92.3–0.99970.66
2-NF2–10–20.7–13.0–0.99570.83
1-NP2–10–20.3–15.2–0.98230.84
3-NFt2–10–16.07.8–0.97661.1

UV detection5-NQ0.5–1010.2–4.00.98290.38
2-NF0.5–108.0–4.20.99370.50
1-NP0.5–1016.0–3.10.99400.24
3-NFt0.5–1021.2–5.00.98820.18

a For electrochemical detection in nA μmol−1 L and for spectrophotometric detection in mA.U. μmol−1 L.
b For electrochemical detection in nA and for spectrophotometric detection in mA.U.