Research Article

Single-Channel Flow Injection Spectrophotometric Determination of Nickel Using Furildioxime in Micellar Solution

Table 1

Selected chemical and FIA parameters obtained with optimization experiments.

ParameterStudied rangeSelected value

Size of sample loop (μL)50–500500
Reagent flow rate (mL min−1)1.0–4.21.2
Length of reaction coil (cm)50–300100
pH3.5–12.09.0
Concentration of reagent (%)0.04–0.250.08
Brij-35 (%)0.01–2.00.01
NH4Cl/NaOH (M)0.04–0.40.2