Development and Validation of an RP-HPLC Method for CB13 Evaluation in Several PLGA Nanoparticle Systems

Table 2

Robustness test of proposed method in terms of recovery (%) and RSD (%) for 500 μg/mL CB13 standard ( 𝑛 = 3 ) (MP: mobile phase).

Changes to original methodTheoretical mean (μg)Peak areaExperimental (μg)Recovery (%)RSD

MP proportion (A : B) 65 : 35500.607787498.33502.83100.450.98
MP proportion (A : B) 75 : 25500.537826118.67507.95101.481.11
MP pH = 2.70500.437821110.33507.29101.370.84
MP pH = 3.10500.507811209505.98101.090.74
Oven temperature −2°C500.467814675506.44101.191.35
Oven temperature +2°C500.307832734508.83101.711.04
Flow rate −0.1 mL/min500.367835018,67509.14101.750.84
Flow rate +0.1 mL/min500.507809528505.75101.080.87