Development and Validation of an RP-HPLC Method for CB13 Evaluation in Several PLGA Nanoparticle Systems
Table 2
Robustness test of proposed method in terms of recovery (%) and RSD (%) for 500 μg/mL CB13 standard () (MP: mobile phase).
| Changes to original method | Theoretical mean (μg) | Peak area | Experimental (μg) | Recovery (%) | RSD |
| MP proportion (A : B) 65 : 35 | 500.60 | 7787498.33 | 502.83 | 100.45 | 0.98 | MP proportion (A : B) 75 : 25 | 500.53 | 7826118.67 | 507.95 | 101.48 | 1.11 | MP pH = 2.70 | 500.43 | 7821110.33 | 507.29 | 101.37 | 0.84 | MP pH = 3.10 | 500.50 | 7811209 | 505.98 | 101.09 | 0.74 | Oven temperature −2°C | 500.46 | 7814675 | 506.44 | 101.19 | 1.35 | Oven temperature +2°C | 500.30 | 7832734 | 508.83 | 101.71 | 1.04 | Flow rate −0.1 mL/min | 500.36 | 7835018,67 | 509.14 | 101.75 | 0.84 | Flow rate +0.1 mL/min | 500.50 | 7809528 | 505.75 | 101.08 | 0.87 |
|
|