Research Article

Ultrasound-Assisted Emulsification Microextraction Based on Solidification Floating Organic Drop Trace Amounts of Manganese Prior to Graphite Furnace Atomic Absorption Spectrometry Determination

Table 1

Temperature program of GFAAS for determination of manganese.

StepsTemperature (°C)Time (S)Argon flow rate (L min−1)Read command

Dry stage855.03.0No
Dry stage9540.03.0No
Dry stage27010.03.0No
Ash stage7005.03.0No
Ash stage7001.03.0No
Gas stop step7002.00.0No
Ramp step and read command24001.10.0Yes
Atomize hold step and read command24002.00.0Yes
Tube clean with maximum gas flow24302.03.0No