Effect of Emitted Wavelength and Light Guide Type on Irradiance Discrepancies in Hand-Held Dental Curing Radiometers
Table 3
Power density calculated from integral value of spectra at 200–800 nm, 380–525 nm, and 430–490 nm (
, mW/cm
2).
| Light-curing unit | Light guide/mode | 200–800 nm | 380–525 nm | 430–490 nm |
| D-Lux 2000 | Straight guide | | | | (100%) | (92.2%) | (64.4%) | Turbo guide | | | | (100%) | (92.1%) | (64.8%) |
| Jetlite 3000 | — | | | | (100%) | (95.0%) | (60.3%) |
| Demi Plus | Straight guide | | | | (100%) | (96.9%) | (93.3%) | Turbo guide | | | | (100%) | (97.5%) | (94.1%) |
| Pencure | — | | | | (100%) | (96.9%) | (89.3%) |
| G-Light Prima | Normal | | | | (100%) | (97.5%) | (90.6%) | F5 | | | | (100%) | (97.1%) | (89.7%) | PH | | | | (100%) | (97.3%) | (74.4%) | PL | | | | (100%) | (82.3%) | (1.8%) |
| Bluephase G2 | High | | | | (100%) | (97.4%) | (79.5%) | Low | | | | (100%) | (97.1%) | (67.1%) |
|
|