Effect of Emitted Wavelength and Light Guide Type on Irradiance Discrepancies in Hand-Held Dental Curing Radiometers
Table 4
Displayed light output (Mean ± S.D.; mW/cm
2) and percentage of the reduction from USR-45 (%) (
).
| | USR-45 (380–525 nm) | Bluephase Meter | Cure Rite | Curing Light Meter 105 | Model 100 Optilux | Model L.E.D. Radiometer |
| D-Lux 2000 | | | | | | | (Straight) | (100%) | (84.9%) | (58.5%) | (42.0%) | (84.6%) | (60.7%) | D-Lux 2000 | | | | | | | (Turbo) | (100%) | (98.5%) | (51.0%) | (51.6%) | (73.4%) | (53.1%) | Jetlite 3000 | | | | | | | (100%) | (75.6%) | (56.2%) | (47.2%) | (71.2%) | (54.5%) | Demi Plus | | | | | | | (Straignt) | (100%) | (84.3%) | (66.3%) | (60.7%) | (97.4%) | (74.0%) | Demi Plus | | | | | | | (Turbo) | (100%) | (90.2%) | (71.6%) | (86.3%) | (51.4%) | (72.4%) | Pencure | | | | | | | (100%) | (70.2%) | (74.2%) | (67.1%) | (80.9%) | (68.8%) | G-Light Prima | | | | | | | (Normal) | (100%) | (82.3%) | (96.7%) | (42.6%) | (48.4%) | (54.6%) | G-Light Prima | | | | | | | (F5 mode) | (100%) | (82.5%) | (61.8%) | (77.2%) | (30.9%) | (60.2%) | G-Light Prima | | | | | | | (PH mode) | (100%) | (70.8%) | (94.7%) | (37.9%) | (47.3%) | (47.8%) | G-Light Prima | | | | | | | (PL mode) | (100%) | (0%) | (0%) | (0%) | (0%) | (0%) | Bluephase G2 | | | | | | | (High) | (100%) | (88.1%) | (58.4%) | (65.1%) | (67.2%) | (69.7%) | Bluephase G2 | | | | | | | (Low) | (100%) | (82.0%) | (49.8%) | (49.1%) | (85.9%) | (67.5%) |
|
|
Within the same row, different letters indicate groups that are statistically different ().
*Achieved to the highest limit of radiometer concerned.
|