Effect of Emitted Wavelength and Light Guide Type on Irradiance Discrepancies in Hand-Held Dental Curing Radiometers

Table 4

Displayed light output (Mean ± S.D.; mW/cm2) and percentage of the reduction from USR-45 (%) ( ).

USR-45
(380–525 nm)
Bluephase MeterCure RiteCuring Light Meter 105Model 100 OptiluxModel L.E.D. Radiometer

D-Lux 2000
(Straight)(100%)(84.9%)(58.5%)(42.0%)(84.6%)(60.7%)
D-Lux 2000
(Turbo)(100%)(98.5%)(51.0%)(51.6%)(73.4%)(53.1%)
Jetlite 3000
(100%)(75.6%)(56.2%)(47.2%)(71.2%)(54.5%)
Demi Plus
(Straignt)(100%)(84.3%)(66.3%)(60.7%)(97.4%)(74.0%)
Demi Plus
(Turbo)(100%)(90.2%)(71.6%)(86.3%)(51.4%)(72.4%)
Pencure
(100%)(70.2%)(74.2%)(67.1%)(80.9%)(68.8%)
G-Light Prima
(Normal)(100%)(82.3%)(96.7%)(42.6%)(48.4%)(54.6%)
G-Light Prima
(F5 mode)(100%)(82.5%)(61.8%)(77.2%)(30.9%)(60.2%)
G-Light Prima
(PH mode)(100%)(70.8%)(94.7%)(37.9%)(47.3%)(47.8%)
G-Light Prima
(PL mode)(100%)(0%)(0%)(0%)(0%)(0%)
Bluephase G2
(High)(100%)(88.1%)(58.4%)(65.1%)(67.2%)(69.7%)
Bluephase G2
(Low)(100%)(82.0%)(49.8%)(49.1%)(85.9%)(67.5%)

Within the same row, different letters indicate groups that are statistically different ( ).
*Achieved to the highest limit of radiometer concerned.