Research Article

Dispersive Liquid-Liquid Microextraction Combined with Ultrahigh Performance Liquid Chromatography/Tandem Mass Spectrometry for Determination of Organophosphate Esters in Aqueous Samples

Table 3

Recoveries and precisions of OPEs.

AnalyteSpiked (µg/L)Recovery (%)RSD (%, )Interday variability (%, )

TEP0.3, 0.6, 3.048.7, 52.8, 58.36.9, 6.2, 6.96.8, 5.6, 5.3
TCEP0.3, 0.6, 3.085.9, 86.8, 91.49.3, 8.1, 5.912.3, 10.2, 5.8
TCPP0.2, 0.4, 2.0107, 94.2, 95.67.3, 5.3, 4.17.0, 5.7, 4.3
TBP0.1, 0.2, 1.090.8, 86.4, 89.43.9, 3.0, 4.32.6, 2.8, 3.0
TBEP0.1, 0.2, 1.0113, 95.6, 1027.4, 3.2, 3.87.5, 5.6, 3.9
TPhP0.1, 0.2, 1.0101, 97.2, 95.86.5, 5.7, 4.54.6, 4.5, 3.8
o-TTP0.1, 0.2, 1.0103, 94.6, 98.86.9, 3.7, 3.23.2, 4.5, 6.4
TEHP0.1, 0.2, 1.0107, 99.8, 1025.0, 3.0, 3.64.3, 3.3, 2.7