Research Article

How Do Metal Ion Levels Change over Time in Hip Resurfacing Patients? A Cohort Study

Table 3

Prospective clinical, radiographic, and biochemical evaluation of nineteen HR patients.

Code at follow-upGenderAge at surg.Status*2-year follow-up 5-year follow-up 9-year follow-up Acetab. Ø mmHead Ø mmCup inclin. angle°HHSUCLA scale
CrCoNiMoCrCoNiMoCrCoNiMo

(1) 3712/4567M52U1.620.510.670.831.170.780.170.9356505892/975/6
(2) 2819/3546/4556F42U1.730.410.690.982.611.440.770.833.812.450.490.8346384090/94/959/9/10
(3) 3561/4564M45U2.472.271.030.831.410.870.390.83625445999/10
(4) 2808/4542M42U1.090.080.690.831.300.530.20.975650321007
(5) 3559/4546M50U1.340.370.970.830.840.360.410.83625448100/989/7
(6) 2817/3567M60U0.820.530.791.040.880.550.730.8360543010010
(7) 2812/3464/4547M55U0.880.230.550.831.160.520.510.960.860.770.30.836054461009/8/7
(8) 3713/4585M61U1.000.830.510.831.010.9990.320.8362544796/1006/10
(9) 2894/4554M42U1.080.110.10.831.060.680.240.83524646100/988
(10) 2830/4583F37U2.080.580.940.831.240.640.790.8354464390/979
(11) 2882/4559F73U7.248.960.481.2328.0332.240.471.5952465983/577
(12) 3664/4568M62U0.910.30.340.830.730.580.280.8358505695/855/6
(13) 2845/4562F42U5.671.670.690.832.301.60.10.8354465584/797
(14) 2844/3460/4544F57U2.360.740.60.831.781.360.940.830.930.930.230.9148424286/90/706/6/3
(15) 2895/3714/4566M24U1.330.450.990.831.600.720.990.831.710.820.420.83584640100/89/957/6/4
(16) 2713/4541M63U0.690.760.340.834.10 2.390.390.93546045100/987
(17) 2816/3568F59B2.991.620.691.32.622.060.64**52504590/1007
(18) 2839/3544/4560††F47U/B2.570.960.630.834.782.890.6**2.981.530.610.83544645100/97/697/7/3
(19) 2810/3558†††M39U/B1.790.630.740.834.212.380.43**605450100/978/9

Bilateral implant from 2 yrs and 5 yrs, ††one and four years after contralateral surgery, †††one year after contralateral surgery, *U: unilateral/B: bilateral, **insufficient serum for Mo.