Research Article

Optimization (Central Composite Design) and Validation of HPLC Method for Investigation of Emtricitabine Loaded Poly(lactic-co-glycolic acid) Nanoparticles: In Vitro Drug Release and In Vivo Pharmacokinetic Studies

Table 3

System suitability parameters.

ParameterCompound
EmtricitabineIS

Retention time (Rt)4.396.58
Tailing factor ( )1.011.13
Injection repeatability (RSD)0.7520.623
Resolution ( )7.32
Capacity factor ( )5.926.30
Theoretical plates (N)49765318
Asymmetry1.231.36

RSD of peak areas of six consecutive injections at a concentration of 2.0 and 0.2 µg/mL of emtricitabine and IS, respectively.
Resolution between emtricitabine and IS.