Research Article

Mineral Element Contents in Commercially Valuable Fish Species in Spain

Table 1

Instrumental analytical conditions for atomic absorption spectrometry (flame and hydride generation technique) and for atomic emission spectrometry (flame technique) of investigated elements.

Wavelength (nm)Slit width (nm)Flow air/acetylene (L/min)Lamp current (%)

FAAS
 Ca422.70.51.2100
 Mg285.20.50.975
 Zn213.90.50.975
FAES
 Na589.00.51.10
 K766.50.51.10

Wavelength (nm)Slit width (nm)Flow argon (L/min)Lamp current (%)

HGAAS
 As193.710.275
CVAAS
 Hg253.70.50.275

FAAS: flame atomic absorption spectrometry; FAES: flame atomic emission spectrometry; HGAAS: hydride generation atomic absorption spectrometry; CVAA: cold vapour atomic spectrometry.