Risk of Fungal Infection to Dental Patients

Table 1

Contamination level (CFU/mL) of water in five pieces of dental equipment obtained in duplicate at four collection points in two dental clinics of the University of Franca in three different sets (with deionized water, with sterile water, and with replaced plastic connections).

Number of samplesHRTSRUCR
ABTotalABTotalABTotalABTotal

SCD1213101112000
SCD2000112112000
SCD322401119120101
SCD411201100002121
SCD50331202111202020
PCD122022202303000
PCD215610118902222
PCD320626412164610055
PCD421021101101000
PCD50000202042603434
Mean8.70 6.605.5010.30
SD10.10 8.666.1112.68
SE6.26 5.373.79 7.86
UL14.9611.979.2918.16
IL 2.44 1.231.71 2.44

SCS1101303202000
SCS2000000000000
SCS3000000000000
SCS4022303000055
SCS503232000011033
PCS1000000404000
PCS2000000000202
PCS3101000000101
PCS4022000202055
PCS5011000000011
Mean3.900.600.901.70
SD9.901.261.372.00
SE6.140.780.851.24
UL10.041.381.752.94
IL0.050.46

SCRC139039000101202
SCRC240347410122022011
SCRC320315100030332133
SCRC40202002525011112
SCRC502020000011112
PCRC123023000112314
PCRC23113210133033011
PCRC31501530300035136
PCRC413435000011112
PCRC50131301818000011
Mean32.204.806.408.40
SD 18.869.0011.4513.80
SE 11.695.587.108.55
UL 43.8910.3813.5016.95
IL 20.51

SCD: Surgical Clinic with deionized water; PCD: Periodontics Clinic with deionized water; SCS: Surgical Clinic with sterile water; PCS: Periodontics Clinic with sterile water; SCRC: Surgical Clinic with replaced plastic connections; PCRC: Periodontics Clinic with replaced plastic connections; SD: standard deviation; SE: standard error; UL: upper limit confidence interval 95%; IL: lower confidence interval 95%; HR: high-revving engine; TS: triple syringes; RU: reservoirs of units (equipment); CR: central reservoir; 1 to 5 refer to the number of pieces of equipment at each collection point (five pieces of equipment for each collection point).