Research Article

Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus

Table 1

Female harlequin duck serum chemistry baseline reference values as determined on paired samples analyzed using the portable VetScan machine and a laboratory analyzer (Monarch).

VetScanMonarch
ParameterUnitsNMedianQuartilesNMedianQuartiles

Glucosemmol/L8818.816.9 –20.48218.417.0–20.2
Potassiummmol/L882.72.4–3.0822.02.0–2.1
Alkaline phosphataseU/L86500291–77082253142–377
Alanine transferaseU/L883526–47821714–22
AmylaseU/L8814301191–163582978827–1169
Total bilirubinμmol/L793.43.4–5.113.4
Calciummmol/L882.32.3–2.4822.32.3–2.5
Cholesterolmmol/L885.64.5–6.5825.34.3–6.2
Creatinineμmol/L8817.717.7–26.58244.235.4–53.0
Total proteing/L883936–41823432–37
Globuling/LNANANA822119–23