The Canine Erythrocyte Sedimentation Rate (ESR): Evaluation of a Point-of-Care Testing Device (MINIPET DIESSE)

Figure 1

The Passing-Bablok regression of group 1 (all blood samples collected). ESR-W, erythrocyte sedimentation rate using the Westergren method; ESR-MP, erythrocyte sedimentation rate using the MINIPET device. Additional information: intercept-systematic difference was 0.33 (95% CI: 0.31 to 0.57); slope-proportional difference was 0.67 (95% CI: 0.42 to 0.98).