Journals
Publish with us
Publishing partnerships
About us
Blog
International Journal of Analytical Chemistry
Journal overview
For authors
For reviewers
For editors
Table of Contents
Special Issues
International Journal of Analytical Chemistry
/
2017
/
Article
/
Tab 4
/
Research Article
GC Method Validation for the Analysis of Menthol in Suppository Pharmaceutical Dosage Form
Table 4
Accuracy results for menthol.
concentration%
Concentration (mg/mL)
Preparation #
Average % recovered
80%
0.48
1
99.5
2
100.4
3
101.3
Average
100.4
% RSD
0.896
100%
0.60
1
101.1
2
98.4
3
99.6
Average
99.7
% RSD
1.370
120%
0.72
1
101.3
2
100.3
3
100.1
Average
100.6
% RSD
0.64
to menthol concentration in standard solution.