Research Article
A Turbidity Test Based Centrifugal Microfluidics Diagnostic System for Simultaneous Detection of HBV, HCV, and CMV
Table 1
Reagent values for each virus.
| | Unit | HBV | HCV | CMV |
| Betaine | M | 1 | 0.8 | 0.8 |
MgSO4 | mM | 4 | 6 | 10 | dNTPs | mM | 1.6 | 1.4 | 1.4 | Bst polymerase | U | 8 | F3/B3 | μM | 0.2 | LF/LB | μM | 0.4 | 1 | 0.8 | FIP/BIP | μM | 1.6 | 2 | 1.6 | Temperature | °C | 63 | 65 | 65 | Time | Minutes | 45 | 60 | 60 | DNA volume | μL | 5 | Sensitivity | Copies | 50 | 60 | 10 |
|
|