Research Article

Comparative Study of the Effect of Structural Parameters on the Flexibility of Endovascular Stent Grafts

Table 5

Luminal reduction rate of commercial stent grafts.

Stent graftsLength of graft (mm)Stent spacing (mm)Stent height (mm)Ratio of stent spacing to stent heightLuminal reduction rate level at 90°

Endurant91.52.18.30.25Low
Zenith Flex99.45.811.70.50High
Zenith LP92.93.48.60.40Medium