Dynamic Contrast-Enhanced Imaging as a Prognostic Tool in Early Diagnosis of Prostate Cancer: Correlation with PSA and Clinical Stage

Table 1

Sequence parameters for 3T multiparametric MRI with the body and spine matrix combination coil system.

SequencePulse sequenceTR (msec)TE (msec)FA (°)FOV (mm)ACQmatrixSlice/gap (mm)

Axial DWI, b = 50, 400, and 800 s/mm2SE-EPI38007790221 × 260102 × 1603.6/0
Axial T2WTSE400010090200 × 200288 × 3203/0.6
Sagittal T2WTSE500010090200 × 200288 × 3203/0.6
Coronal T2WTSE500010090200 × 200288 × 3203/0.6
Axial 3DFLASH GRE4.91.72 and 13260 × 260138 × 1923/0
Axial 3D DCEFLASH GRE4.91.712260 × 260138 × 1923.6/0

SE, spin echo; EPI, echo planar imaging; TSE, turbo spin echo; FLASH, fast low angle shot; GRE, gradient recalled echo; TR, repetition time; TE, echo time; FA, flip angle; ACQ matrix, acquisition matrix. Sequence for the measurement of T1 relaxation time.