Case Report | Open Access

Volume 2010 |Article ID 741751 | https://doi.org/10.4061/2010/741751

Venkata M. Alla, Yeruva M. Reddy, William Abide, Tom Hee, Claire Hunter, "Delayed Lead Perforation: Can We Ever Let the Guard Down?", Cardiology Research and Practice, vol. 2010, Article ID 741751, 4 pages, 2010. https://doi.org/10.4061/2010/741751

Delayed Lead Perforation: Can We Ever Let the Guard Down?

Academic Editor: Brian Olshansky
Received05 Apr 2010
Revised26 Jun 2010
Accepted27 Jun 2010
Published25 Jul 2010

Abstract

Lead perforation is a major complication of cardiac rhythm management devices (CRMD), occurring in about 1%. While most lead perforations occur early, numerous instances of delayed lead perforation (occurring 30 days after implantation) have been reported in the last few years. Only about 40 such cases have been published, with the majority occurring 1 year after implantation. Herein, we describe the case of an 84-year-old female who presented with recurrent syncope and was diagnosed to have delayed pacemaker lead perforation 4.8 years after implantation. Through this report, we intend to highlight the increasing use of CRMD in elderly patients, and the lifelong risk of complications with these devices. Presentation can be atypical and a high index of suspicion is necessary for diagnosis.

1. Introduction

Increasing burden of cardiovascular disease and broadening indications have led to a significant increase in the implantation of cardiac rhythm management devices (CRMDs) [13]. Approximately 70% of CRMD recipients are 65 years of age and 20%–35% are older than 80 years [14]. Cardiac perforation is a major complication of CRMD implantation and can be acute, subacute, or delayed. Herein, we describe a case of delayed pacemaker lead perforation (DLP) occurring 4.8 years after implantation. Through this report, we emphasize the need to consider device related complications in the differential diagnoses of patients with CRMD, even years after implantation.

2. Case Summary

An 84-year-old female presented with two episodes of transient loss of consciousness over 24 hours. She was initially evaluated in the emergency room of an outlying hospital where physical examination, electrocardiogram (ECG), and chest X-ray were unremarkable. Subsequently, she was transferred to our hospital for further evaluation. She denied chest pain, palpitations, dyspnea, or orthostatic dizziness. She was seated at the time of the index events, had no premonitory symptoms, and regained consciousness within a few minutes. Past medical history was significant for rheumatoid arthritis, paroxysmal atrial fibrillation, and symptomatic sinus pauses requiring a dual chamber pacemaker placement. The pacemaker (Medtronic, KDR901 Kappa DR; leads: Medtronic 5076 CapSure Fix Novus) was implanted in December, 2004 with the right ventricular lead in the apex and right atrial lead in the appendage. Her medications included warfarin, metoprolol, prednisone, and amiodarone. Pulse rate was 65/minute, blood pressure was 96/58 mmHg, and physical examination was significant for prominent jugular venous pulsations, and distant heart sounds. Metabolic panel and blood counts were normal and INR was therapeutic at 2.1. ECG revealed normal sinus rhythm without any abnormalities, chest X-ray demonstrated mild cardiomegaly with stable atrial/ventricular lead position and computed tomography scan of the head was unremarkable. Transthoracic echocardiogram revealed a large pericardial effusion with early echocardiographic signs of cardiac tamponade (Figure 1). Pacemaker evaluation revealed normal function; lead impedance, sensing and pacing thresholds were stable compared to evaluation 3 months prior. Previous echocardiogram, also done 3 months prior was unremarkable except for mild left ventricular hypertrophy. She underwent emergent surgical pericardiocentesis and about 600 mL of bloody fluid was drained. During the surgery, a defect in the right ventricular myocardium was visualized and repaired with sutures. The ventricular lead was in close proximity but there was no definite protrusion of the tip through the defect. A pericardial window was created and the right ventricular lead was successfully repositioned under transesophageal echo guidance. Pericardial fluid cultures were sterile, cytology was negative for malignant cells, and pericardial biopsy was normal. Further hospital stay was uneventful and she was discharged on her home medication regimen. She has had no recurrent symptoms or pericardial effusion at 1 year followup.

3. Discussion

The frequency of lead perforation varies between 0.1%–0.8% for pacemaker leads and 0.6%–5.2% for defibrillators [57]. Lead perforation is considered to be acute when it occurs within 5–7 days after implantation; subacute when it occurs between 7–30 days; delayed when it occurs beyond 30 days after implantation. While most lead perforations occur early, DLP has been increasingly reported in the last few years. We identified a total of 38 patients with DLP in our review of literature (Table 1). The vast majority of these occurred between 6 weeks and 1 year from implantation; 7/38 patients had DLP beyond one year and only one patient had this complication 3 years after implantation. Our patient presented close to 5 years after implantation, which is unusually long and underscores the potential lifelong risk.


AuthorPublicationNumber of patientsAge in yearsTime from implantation to perforationComments

Ahmet AkyolPacing Clin Electrophysiol 2005; 28:350–3511246 months
Khan MNPacing Clin Electrophysiol 2005; 28:251–253326, 71, 816–10 months
Velavan PHeart 2003; 89:3641641 month
Satpathy RPacing Clin Electrophysiol 2008; 31:10–1217210 months
Haq SAAngiology 2008; 59:61918616 months
Ellenbogen KAPacing Clin Electrophysiol 2002; 25:1155–1158373, 72, 4230 days3/5 cases reported here were late perforations others being sub acute.
Kautzner JPacing Clin Electrophysiol 2001; 24:116–11813623 monthsSymptoms of pericarditis without perforation 4 months after implantation.
Polin GMAm J Cardiol. 2006 15; 98:223–5538, 55, 79, 85, 886 weeks–3years
Laborderie JAm J Cardiol 2008; 102:1352–1355835, 50, 40, 53, 64, 78, 81, 841–3.5 months8/11 cases reported here were late perforations.
Fisher JDPacing Clin Electrophysiol 2008; 31:7–917138 days
Krivan LPacing Clin Electrophysiol 2008; 31:3–61471 month1/2 cases reported here was late perforation the other was acute.
Lloyd MSPacing Clin Electrophysiol 2008; 31:784–7851686 weeks1/3 cases reported here was late perforation others being sub acute.
Suri RHeart Rhythm. 2007 Sep; 4(9):1248–91NA6 weeks1/5 cases reported here was late perforation others being sub acute.
Sadamatsu KJ Cardiol. 2009; 53(1):150–3.1NA9 monthsasymptomatic
Kanoh MKyobu Geka. 1994; 47:730–11803 monthsasymptomatic
Park REPacing Clin Electrophysiol 2008; 31:785–7861726 weeks
Singhal SCirculation. 2007; 115:e391-21507 yearsPerforated through the rib.
Wiegand UKPacing Clin Electrophysiol. 2003; 26(10):1961-91NA3 years1 case of delayed perforation in 116 implanted leads.
Sanoussi APacing Clin Electrophysiol. 2005; 28:723–7251791 month
Tziakas DEuropace 2009; 11:968–969.1845 weeks
Celik TEuropace 2009; 11:963–965.273, 658 months, 2 years
Tavernier REuropace. 2009; 11:966–71741 month

Though the lead was not visualized in the pericardial space we believe that DLP is the most likely explanation in our patient. The lead was noted to be in close proximity to the defect in the RV myocardium, effusion was bloody, and there was no alternative etiology for myocardial perforation or pericardial effusion. However, the exact timing of lead perforation is unclear. Though chronic perforation cannot definitively be excluded, a normal echocardiogram 3 months prior suggests that this was more recent. Presentation of DLP is variable ranging from minimal or no symptoms to cardiac tamponade and sudden death. Several known risk factors for acute lead perforation exist; these include patient characteristics like advanced age, female gender; hardware factors like defibrillator versus pacemaker leads, smaller lead diameter, increased lead stiffness, active versus passive

fixation tips; and miscellaneous factors like use of steroids in the perioperative period [8]. However, the role of these factors in DLP has not been clearly established.

In summary, patients continue to be at risk for potentially grave complications like lead perforation many years after implantation of CRMD. Symptoms can be subtle or atypical especially in those with advanced age and a high index of suspicion is necessary for timely diagnosis.

Conflict of Interest/Disclosures

None of the authors have any potential conflict of interest or relevant disclosures.

Authors’ Contribution

All the authors had access to the data and contributed to the content and writing of the paper. The paper is entirely original. It has not been previously published nor is under consideration for publication elsewhere. All authors hereby transfer complete copyright ownership of this paper to the Cardiology research and practice journal in the event this work is published. All listed authors authorize the corresponding author to commit to copyright transfer on their behalf.

References

  1. Z. Goldberger and R. Lampert, “Implantable cardioverter-defibrillators: expanding indications and technologies,” Journal of the American Medical Association, vol. 295, no. 7, pp. 809–818, 2006. View at: Publisher Site | Google Scholar
  2. C. Zhan, W. B. Baine, A. Sedrakyan, and C. Steiner, “Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis,” Journal of General Internal Medicine, vol. 23, supplement 1, pp. 13–19, 2008. View at: Publisher Site | Google Scholar
  3. D. Z. Uslan, I. M. Tleyjeh, L. M. Baddour, P. A. Friedman, S. M. Jenkins, J. L. St Sauver, and D. L. Hayes, “Temporal trends in permanent pacemaker implantation: a population-based study,” American Heart Journal, vol. 155, no. 5, pp. 896–903, 2008. View at: Publisher Site | Google Scholar
  4. G. Lin, R. A. Meverden, D. O. Hodge, D. Z. Uslan, D. L. Hayes, and P. A. Brady, “Age and gender trends in implantable cardioverter defibrillator utilization: a population based study,” Journal of Interventional Cardiac Electrophysiology, vol. 22, no. 1, pp. 65–70, 2008. View at: Publisher Site | Google Scholar
  5. K. A. Ellenbogen, M. A. Wood, and R. K. Shepard, “Delayed complications following pacemaker implantation,” Pacing and Clinical Electrophysiology, vol. 25, no. 8, pp. 1155–1158, 2002. View at: Google Scholar
  6. M. N. Khan, G. Joseph, Y. Khaykin, K. M. Ziada, and B. L. Wilkoff, “Delayed lead perforation: a disturbing trend,” Pacing and Clinical Electrophysiology, vol. 28, no. 3, pp. 251–253, 2005. View at: Publisher Site | Google Scholar
  7. M. D. Carlson, R. A. Freedman, and P. A. Levine, “Lead perforation: incidence in registries,” Pacing and Clinical Electrophysiology, vol. 31, no. 1, pp. 13–15, 2008. View at: Publisher Site | Google Scholar
  8. S. Mahapatra, K. A. Bybee, T. J. Bunch, R. E. Espinosa, L. J. Sinak, M. D. McGoon, and D. L. Hayes, “Incidence and predictors of cardiac perforation after permanent pacemaker placement,” Heart Rhythm, vol. 2, no. 9, pp. 907–911, 2005. View at: Publisher Site | Google Scholar

Copyright © 2010 Venkata M. Alla et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder
Views1053
Downloads592
Citations

Related articles

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.