Abstract

Lead perforation is a major complication of cardiac rhythm management devices (CRMD), occurring in about 1%. While most lead perforations occur early, numerous instances of delayed lead perforation (occurring 30 days after implantation) have been reported in the last few years. Only about 40 such cases have been published, with the majority occurring 1 year after implantation. Herein, we describe the case of an 84-year-old female who presented with recurrent syncope and was diagnosed to have delayed pacemaker lead perforation 4.8 years after implantation. Through this report, we intend to highlight the increasing use of CRMD in elderly patients, and the lifelong risk of complications with these devices. Presentation can be atypical and a high index of suspicion is necessary for diagnosis.

1. Introduction

Increasing burden of cardiovascular disease and broadening indications have led to a significant increase in the implantation of cardiac rhythm management devices (CRMDs) [13]. Approximately 70% of CRMD recipients are 65 years of age and 20%–35% are older than 80 years [14]. Cardiac perforation is a major complication of CRMD implantation and can be acute, subacute, or delayed. Herein, we describe a case of delayed pacemaker lead perforation (DLP) occurring 4.8 years after implantation. Through this report, we emphasize the need to consider device related complications in the differential diagnoses of patients with CRMD, even years after implantation.

2. Case Summary

An 84-year-old female presented with two episodes of transient loss of consciousness over 24 hours. She was initially evaluated in the emergency room of an outlying hospital where physical examination, electrocardiogram (ECG), and chest X-ray were unremarkable. Subsequently, she was transferred to our hospital for further evaluation. She denied chest pain, palpitations, dyspnea, or orthostatic dizziness. She was seated at the time of the index events, had no premonitory symptoms, and regained consciousness within a few minutes. Past medical history was significant for rheumatoid arthritis, paroxysmal atrial fibrillation, and symptomatic sinus pauses requiring a dual chamber pacemaker placement. The pacemaker (Medtronic, KDR901 Kappa DR; leads: Medtronic 5076 CapSure Fix Novus) was implanted in December, 2004 with the right ventricular lead in the apex and right atrial lead in the appendage. Her medications included warfarin, metoprolol, prednisone, and amiodarone. Pulse rate was 65/minute, blood pressure was 96/58 mmHg, and physical examination was significant for prominent jugular venous pulsations, and distant heart sounds. Metabolic panel and blood counts were normal and INR was therapeutic at 2.1. ECG revealed normal sinus rhythm without any abnormalities, chest X-ray demonstrated mild cardiomegaly with stable atrial/ventricular lead position and computed tomography scan of the head was unremarkable. Transthoracic echocardiogram revealed a large pericardial effusion with early echocardiographic signs of cardiac tamponade (Figure 1). Pacemaker evaluation revealed normal function; lead impedance, sensing and pacing thresholds were stable compared to evaluation 3 months prior. Previous echocardiogram, also done 3 months prior was unremarkable except for mild left ventricular hypertrophy. She underwent emergent surgical pericardiocentesis and about 600 mL of bloody fluid was drained. During the surgery, a defect in the right ventricular myocardium was visualized and repaired with sutures. The ventricular lead was in close proximity but there was no definite protrusion of the tip through the defect. A pericardial window was created and the right ventricular lead was successfully repositioned under transesophageal echo guidance. Pericardial fluid cultures were sterile, cytology was negative for malignant cells, and pericardial biopsy was normal. Further hospital stay was uneventful and she was discharged on her home medication regimen. She has had no recurrent symptoms or pericardial effusion at 1 year followup.

3. Discussion

The frequency of lead perforation varies between 0.1%–0.8% for pacemaker leads and 0.6%–5.2% for defibrillators [57]. Lead perforation is considered to be acute when it occurs within 5–7 days after implantation; subacute when it occurs between 7–30 days; delayed when it occurs beyond 30 days after implantation. While most lead perforations occur early, DLP has been increasingly reported in the last few years. We identified a total of 38 patients with DLP in our review of literature (Table 1). The vast majority of these occurred between 6 weeks and 1 year from implantation; 7/38 patients had DLP beyond one year and only one patient had this complication 3 years after implantation. Our patient presented close to 5 years after implantation, which is unusually long and underscores the potential lifelong risk.

Though the lead was not visualized in the pericardial space we believe that DLP is the most likely explanation in our patient. The lead was noted to be in close proximity to the defect in the RV myocardium, effusion was bloody, and there was no alternative etiology for myocardial perforation or pericardial effusion. However, the exact timing of lead perforation is unclear. Though chronic perforation cannot definitively be excluded, a normal echocardiogram 3 months prior suggests that this was more recent. Presentation of DLP is variable ranging from minimal or no symptoms to cardiac tamponade and sudden death. Several known risk factors for acute lead perforation exist; these include patient characteristics like advanced age, female gender; hardware factors like defibrillator versus pacemaker leads, smaller lead diameter, increased lead stiffness, active versus passive

fixation tips; and miscellaneous factors like use of steroids in the perioperative period [8]. However, the role of these factors in DLP has not been clearly established.

In summary, patients continue to be at risk for potentially grave complications like lead perforation many years after implantation of CRMD. Symptoms can be subtle or atypical especially in those with advanced age and a high index of suspicion is necessary for timely diagnosis.

Conflict of Interest/Disclosures

None of the authors have any potential conflict of interest or relevant disclosures.

Authors’ Contribution

All the authors had access to the data and contributed to the content and writing of the paper. The paper is entirely original. It has not been previously published nor is under consideration for publication elsewhere. All authors hereby transfer complete copyright ownership of this paper to the Cardiology research and practice journal in the event this work is published. All listed authors authorize the corresponding author to commit to copyright transfer on their behalf.