Simple Quantification of Pentosidine in Human Urine and Plasma by High-Performance Liquid Chromatography

Table 1

Mean response factors, coefficients of variation (CV), accuracy, and recovery of pentosidine at various concentrations in human urine.

Concentration (nM)Response factor ± SD (% CV) (%) (%)

181986 ± 7746 (9.45)99.6112
273662 ± 884 (1.20)89.597.2
380179 ± 3599 (4.49)97.4109
483676 ± 5233 (6.25)102111
587057 ± 1527 (1.75)106114
1083147 ± 1457 (1.75)101112
2084339 ± 2234 (2.65)102112
5084359 ± 1510 (1.79)103105

Measurements were conducted three times in separate days; the response factor was calculated by subtracting nonspiked response factor from the spiked standard response factor followed by dividing by the spiked concentration; data are expressed as mean response factor ± standard deviation (SD) ; values in bracket are intraday coefficients of variation (CV). ; recovery compared with water.