Abstract

Polymers have been used in different areas. Recently, polymeric material is favored in analytical area due to its high performance and high consistency, which was used in sample pretreatment in this study. Xylazine poisoning is often seen in body fluid samples obtained from various accidents or suicides. However, the content of xylazine is difficult to detect precisely due to matrix effect in testing practices. In this paper, a method application for phospholipid adsorption polymeric materials to determine xylazine in blood and urine samples was proposed, developed, and validated. Compared with existing method, this method using polymeric pretreatment has a wider linear range of 2.0–2000.0 ng/mL for xylazine and its metabolite 2,6-dimethylaniline in both blood and urine and lower detection limits of 0.3 ng/mL for 2,6-dimethylaniline and xylazine in blood and 0.2 ng/mL for 2,6-dimethylaniline and xylazine in urine. Therefore, this method is suggested to be applied in testing practices by academic groups and commercial organizations.

1. Introduction

The variety of polymers gives different research area different opportunities to achieve different functions. In the recent decades, natural polymers have been modified through grafting to conquer their born defects, including the poor solubility and poor mechanical properties [14]. Meanwhile, synthetic polymers have been engineered so much to achieve new functions, such as long-term antimicrobial polymers [5]. More importantly, polymers have been regarded as the most important component in composite material towards stronger and smarter materials [69] and more precise and more sensitive extraction materials [10, 11]. Especially in the analytical field, polymers play an important role in separation and purification when dealing with novel materials [12, 13] and complicated biosamples [14]. Blood is a complex sample that contains plasma, proteins, and other trace interferents that can often cause matrix effect when using electrospray ionization (ESI) technology. Therefore, the blood sample preparation is a very important step in the analysis process. In recent years, several new methods have been developed and applied to blood sample preparation, for example, protein and phospholipid removal plate; it has been used for the targeted removal of phospholipids from blood samples. The protocol can remove interferents in the blood as much as possible, especially proteins and phospholipids, resulting in lowered matrix effect and enhanced recovery. The operation of the protocol is simple and convenient, which only needs to conduct the protein precipitation inside the plate and then push samples to flow through the absorber. The resultant filtrate can be used directly in LC-MS detection.

Xylazine has been suggested to be used on various kinds of animals, including dogs, cats, horses, and deer, for antianxiety, sedative, and analgetic purposes [2]. The chemical structure of xylazine resembles that of phenothiazines, tricyclic antidepressants, and clonidine. Xylazine is a highly effective α-2 adrenergic stimulant, which can stimulate α-2 receptors at the central nervous system. When those α-2 receptors are stimulated, the central nervous system reduces the release of norepinephrine and dopamine, resulting in the effects of tranquilizer, painkiller, and muscle relaxant. Xylazine can also affect the cholinergic neuronal system, α-1 adrenergic receptor, and histamine system [15]. It has been reported that xylazine has been abused for its anesthetic effect either purposely or accidentally in intoxication and criminal cases [16, 17]. Human bodies will show symptoms such as depressed central nervous system, depressed respiratory system, bradycardia, or low blood pressure when intoxicated with xylazine. The main xylazine biotransformation pathway is most likely thiazine ring break-down, and the main product is most likely 2,6-dimethylaniline [15, 18]; the structure of xylazine and 2,6-dimethylaniline was showed in Figure 1. Due to the qualitative detection ability, chromatography-mass spectrometry has been widely used in biological samples [1922]. The cleanup and target extraction of blood and urine samples have been achieved by different sample preparation methods, such as liquid–liquid extraction (LLE) and solid phase extraction (SPE) [11, 22, 23]. Currently, xylazine detection is usually chromatography or chromatography-mass spectrometry methods. The samples are usually animal-derived foods such as honey and milk [24, 25]. For human intoxications, the detection target is xylazine in blood [26, 27].

To validate this method and to create a reliable sample preparation method quantitatively to determine the xylazine and its metabolite in body fluids, in this paper, phospholipid removal technology was applied in the detection of xylazine and 2,6-dimethylaniline in body fluid sample. At the same time, the results were verified through a comparison with LLE and SPE.

2. Material and Methods

2.1. Chemicals and Reagents

Acetonitrile, methanol, diethyl ether, n-hexane, ethyl acetate, dichloromethane (LC/MS grade, Fisher Scientific, USA), and formic acid (chromatography grade, ROE, USA) were used as received. Water was purified using a Milli-Q water purification system (Millipore, Billerica, MA, USA), Oasis MCX® (30 mg/1 mL; Waters), Cleanert® SLE-AQ (30 mg/1 mL, Bonna-Agela Technologies, China), Ostro® 96 wells (30 mg/1 mL; Waters), xylazine (high purity, 98.0%, J&K Chemical, China), and 2,6-dimethylaniline (high purity, 98.0%, J&K Chemical, China) were used as received, blank blood was supplied by Chongqing city blood center, and blank urine was supplied by healthy laboratory researchers.

2.2. Preparation of Standards and QC Samples

100 mg/L stock solutions of xylazine and 2,6-dimethylaniline were prepared by dissolving 1.0 mg chemicals in 10 mL acetonitrile. After preparation, stock solutions were stored at −20°C until use. Working solution mixtures were prepared by serial dilution of the stock solutions with acetonitrile. Matrix-matched calibration standards and QC samples are done by spiking whole blood and urine with the appropriate volume of the mixed working solution. The concentrations of the matrix-matched calibration standards were 2, 5, 10, 50, 100, 500, and 2000 ng/mL and three QC samples at 5, 50, and 500 ng/mL were prepared for whole blood and urine.

2.3. Instrumentation

LC-MS/MS was performed using a Shimadzu 20A series liquid chromatograph (Shimadzu Corp., Japan) equipped with AB2000 triple quadrupole mass spectrometry (AB Sciex Pte. Ltd., USA). Chromatographic separation was achieved on a Waters Atlantis dC18 (150 mm × 3.9 mm, 5 μm) LC column; temperature: 40°C, sample size: 20 μL, and flow rate: 0.5 mL/min. Mobile phase A is an aqueous solution containing 0.1% formic acid. Mobile phase B is acetonitrile solution containing 0.1% formic acid; the mobile phase is programmed to run in the following sequence: 0-1 min A content is 90%, 2-3 min A content is decreased from 90% to 5%, 3–6 min A content is 5%, 6-7 min A content is increased from 5% back to 90%, and 7-8 min A content is 90%.

The electrospray ionization (ESI) conditions for analyte were as follows: nitrogen was used as nebulizing gas (GS 1, 65 psi), turbo spray gas (GS 2, 65 psi), and curtain gas (20 psi). The collision-activated dissociation (CAD) was set to medium level, while the ion spray voltage was at 4000 V. The dwell time was set at 0.1 seconds, and the Mass Spectrum scanned in positive ion mode.

Simultaneous detection of all glycerophosphocholines (GPCho’s) using Mass Spectrum detection was based on Little et al. [27]. The Mass Spectrum parameters for AB2000 triple quadrupole mass spectrometry have been optimized by employing a declustering potential, entrance potential, collision energy, and collision cell exit potential of 180, 10, 7, and 5 V, respectively. The mass transition of    was monitored in the positive ion electrospray mode.

2.4. Sample Preparation
2.4.1. SPE

Solid phase extraction was described in the previous publications [11, 28]. A 100 μL sample was mixed with 200 μL of 2% formic acid and then was vortexed for 10 s and subjected to SPE. The sample mixture was extracted by Oasis MCX  SPE cartridge. The cartridge was activated subsequently by 1 mL methanol and 1 mL 2% formic acid acidified water right before use. Cartridges were eluted with 2 mL of 2% ammonia in methanol and then dried under a stream of heated nitrogen at 40°C. Finally, the residues were reconstituted in 100 μL initial mobile phase solution. A 20 μL solution was injected into the LC-MS/MS system for analysis.

2.4.2. SLE

A 100 μL sample was mixed with 200 μL deionized water and then was loaded onto a Cleanert SLE-AQ cartridge. A minimum positive pressure was applied to facilitate the sample absorption into the cartridge in less than 10 s. After the analytes were allowed to equilibrate with the sorbent for 10 min, the analytes were eluted with 1.5 mL of extraction solvent each time for three times. The extraction solvent was eluted by applying a slightly positive pressure from the top and then evaporated to dryness under a stream of heated nitrogen at 40°C. Finally, the residues were reconstituted in 100 μL initial mobile phase solution. A 20 μL solution was injected into the LC-MS/MS system for analysis.

2.4.3. Protein and Phospholipid Removal Cartridge

A volume of 300 μL of acetonitrile was transferred to an Ostro 96-well cartridge and then 100 μL plasma was added to the cartridge. The Ostro 96-well cartridge was then manually closed and vortexed. After short moderate mixing, the cartridge was applied to positive pressure and the filtrate was collected into a plastic tube (2 mL Axygen). At this point, the filtrate was ready for immediate LC-MS/MS analysis, but to make this method comparable with the other methods with higher sensitivity, it was concentrated by drying under a nitrogen stream at 40°C and reconstitution in 100 μL of initial mobile phase solution. A 20 μL solution was injected into the LC-MS/MS system for analysis.

2.5. Method Validation

Matrix-matched calibration curves were constructed by plotting the peak area () versus the nominal concentration of the calibration standards (). The linearity of calibration curves was evaluated by fitting calibration curves by weighted least-squares linear regression with a weighting factor of . Selectivity was investigated by averaging results from five blank samples. The limit of detection (LOD) was set at the lowest concentration yielding at least three signal-to-noise () ratios with an acceptable chromatographic peak shape. The limit of quantification (LOQ) was the lowest concentration with an ratio greater than 10.

The recovery and matrix effect were evaluated by repeating 6 analyses at 3 QC samples (5, 50, and 500 ng/mL). The recovery was calculated by dividing the average peak area of the analyte spiked before extraction by the average peak area of the analyte spiked after extraction. The matrix effect was estimated by dividing the average peak area of the analyte spiked after extraction by average peak area derived from the neat standards.

Inaccuracy was determined as the percentage deviation of the average of the results from the corresponding nominal value. Imprecision was expressed as the percent relative standard deviation (% RSD). Intraday precision assays were determined by analyzing QC samples (5, 50, and 500 ng/mL) on the same day ().

3. Results and Discussion

3.1. Optimization of SPE Method

In current studies, three different extraction methods, SPE, SLE, and protein and phospholipid removal, were chosen for comparison. SPE with Oasis MCX polymeric sorbent cartridge, SLE with Cleanert SLE-AQ cartridge, and protein and phospholipid removal cartridge with Ostro 96-well cartridge were investigated during the method development. The extraction efficiencies of these three different extraction methods were evaluated by extracting human blood with 100 ng/mL target addition. As shown in Figure 2, the extraction efficiencies for xylazine with SLE were improved significantly when using ethyl acetate as an eluting solution; the recovery increased from 20% to 95%. However, all five eluting solutions used were not able to extract 2,6-dimethylaniline effectively; the recovery was only from 7% to 75%. The extraction efficiencies for xylazine and 2,6-dimethylaniline with SPE according to the literature [17] were 81% and 75%. In contrast, the extraction efficiencies for xylazine and 2,6-dimethylaniline with Ostro 96-well cartridge are 106% and 103%, respectively. It is obvious to draw the conclusion that Ostro 96-well cartridge has the highest extraction efficiency based on the above comparison. Therefore, Ostro 96-well cartridge was employed to treat the sample. Ostro 96-well protein and phospholipid removal cartridge facilitated protein precipitation technology for the targeted removal of phospholipids from plasma and serum samples, which was fast, simple, and less time and sample consuming and could significantly reduce the matrix effect. According to the results obtained by Bruce et al. [29], acetonitrile has better protein removal efficiency than methanol in blood samples. The addition of 0.1% formic acid to acetonitrile is also beneficial for the protein removal. The parameters evaluated in method development need to be optimized for the volume ratio of acetonitrile to blood. Four different ratios of organic solvent to blood were tested: 1 : 1, 2 : 1, 3 : 1, and 4 : 1 (v/v). 100 μL of blood was used as a blood sample. When the acetonitrile to blood ratio was 1 : 1, the sample solution is thickened, but there is no protein precipitation. When the ratio was increased to 2 : 1, the protein recoveries were 87% and 81% for xylazine and 2,6-dimethylaniline, respectively, suggesting the proteins were not precipitated completely. When the acetonitrile to blood ratio was further increased to 3 : 1 or 4 : 1, the recoveries for both xylazine and 2,6-dimethylaniline were close to 100%. Taking the sensitivity into account, the dilution factor of samples should be as little as possible. Thus, a ratio of 3 : 1 was used.

Glycerophosphocholines (GPCho’s) are known to cause liquid chromatography-mass spectrometry/mass spectrometry (LC-MS/MS) matrix ionization effects during the analysis of blood samples. To evaluate the phospholipid removal performance of three methods, in source collision, induced dissociation [30] was employed to monitor the major GPCho’s causing matrix effect. Blood samples pretreated by the proposed three methods and their 184.2-184.2 ion couples were monitored by MRM method under the same LC condition. The resultant profile is shown in Figure 3. It is apparent that the phospholipid profiles after different pretreatments differ significantly. As indicated in retention time 2–4 min, phospholipids removal by Ostro cartridge is more efficient than the other two methods, although phospholipids are not completely removed. Also, there are no phospholipid peaks in the mass spectra in the 6.5–7.5 min range of blood sample pretreated by Ostro cartridge method. In contrast, blood samples pretreated by SPE and SLE methods show apparent phospholipid peaks. Thus, it can be concluded that the phospholipid removal efficiency is in the following order: Ostro cartridge > SPE > SLE. Ostro cartridge has the best phospholipid removal and lowest matrix effect, which significantly enhances the target recovery and accuracy. Also, it can provide other potential benefits such as longer LC column lifetime due to the lowered phospholipid content. Regarding pretreatment time, Ostro cartridge method requires the shortest time among the three since it only needs filtration, while SPE requires multiple step procedures to activate and rinse the extraction column and SLE requires a waiting time of 5–10 minutes to reach equilibrium. In other words, Ostro cartridge method is the best among the three for high-throughput sample processing.

3.2. Optimization of Chromatographic Separation

Mobile phase composition and addition would affect the targets separation and ionization efficiencies. Two organic solvents, methanol and acetonitrile, as mobile phases were investigated for the effect of mobile phases on the retention time and peak shapes. When methanol was used as mobile phase, the peak showed an apparent tail and its base had been widening. It had longer retention time too. In contrast, the peak shape improved, and retention time is shorter when acetonitrile was used as mobile phase, at the expense of baseline noise. The amine groups on xylazine and 2,6-dimethylaniline could facilitate ionization under acidic conditions. 0.05%, 0.10%, and 0.15% formic acid were added into mobile phase to study the sensitivity of target compounds at various acidities. The results (not shown) suggested the highest sensitivity at 0.10% formic acid. There was no significant improvement in sensitivity when increasing formic acid concentration from 0.10% to 0.15%. The addition of formic acid into acetonitrile mobile phase greatly improved the sensitivity of both xylazine and 2,6-dimethylaniline, which was more than enough to offset the baseline noise. Thus, the mobile phase composition was selected to be 0.10% formic acid in acetonitrile and water. Such mobile phase, with gradient elution, could be helpful to improve the peak shape and enhance ionization efficiencies.

3.3. Optimization of MS Detection

The amine functional groups on both target compounds could be easily protonated at lower pH environments, making the two target compounds suitable to be analyzed under positive ion mode. 2,6-Dimethylaniline and xylazine protonated molecules were observed in the full spectra MS scan under positive mode at 122 and 221, respectively, when tuned with a single standard solution. After the affirmation of precursor ions, more than two products should be chosen when using LC-MS analysis.

Then MS/MS scan was performed to each precursor ion; the major fragments of 2,6-dimethylaniline were at 105 and 77, which were selected to construct MRM transitions with precursor ion at 122 (Figure 4(a)). Five major fragments of xylazine were observed at 164, 147, 121, 105, and 89 (Figure 4(b)) [25]. Last but not least, MRM transitions with 164 and 89 for 211 were selected to monitor in the determination of xylazine. The addition of formic acid into mobile phase could enhance ionization efficiencies, target responses, and sensitivities. The optimized parameters, including DP, CE, and CXP, were listed in Table 1.

3.4. Method Validation

In the MRM mode, the high selectivity for the determination of xylazine and 2,6-dimethylaniline in blood and urine samples was found. The typical MS/MS spectra of spiked blood and urine samples and the blank blood and urine samples are depicted in Figure 5. No additional peaks due to endogenous substances that could have interfered with the detection of xylazine and 2,6-dimethylaniline were observed. The LODs were 0.3 and 0.2 ng/mL for xylazine and 2,6-dimethylaniline in blood and urine. The LOQs were 1.0 and 0.6 ng/mL for xylazine and 2,6-dimethylaniline in blood and urine in Table 2. Standard calibration curves were linear over the curve range between 2 and 2000 ng/mL in blood and urine for both xylazine and 2,6-dimethylaniline; the linear regression equations were in Table 2.

Recovery precision and matrix effect (ME) were repeating analysis for each QC sample 6 times. The QC samples were solutions containing 5, 50, and 500 ng/mL target compounds. The results are summarized in Table 2. The extraction recovery rates were consistent and greater than 93.8%, which indicated that protein and phospholipid removal cartridge was optimal for xylazine and 2,6-dimethylaniline. The precision, assessed by the relative standard deviation (RSD) for the replicate analysis, was typically less than 9%.

Since the selectivity of the HPLC-MS/MS method could be affected by matrix effects by the sample matrix and interferences from compound metabolites, matrix effects are investigated to develop a more specific and reliable method. The matrix effect ranged from 101.1 to 108.2% for both analytes. This result suggested that ion suppression or enhancement from the blood and urine matrix was negligible under the current treatment method. With a low matrix effect and consistent and reproducible recovery, this assay has proved to be reliable for bioanalysis.

3.5. Application to Actual Samples

A 36-year-old woman ingested a certain amount of xylazine after a quarrel with her husband and became unconscious. She was immediately sent to the hospital when her family discovered and was found dead at the time of arriving at hospital. A heart blood sample was collected from the subject for toxicological analysis; xylazine and 2,6-dimethylaniline were confirmed and quantified at 267 ng/mL and 451 ng/mL in the whole blood sample.

4. Conclusions

In this study, an HPLC-MS/MS method for the quantification of xylazine and 2,6-dimethylaniline in blood and urine samples was developed and validated. A rapid, sensitive, specific, selective, and reproducible assay was developed. Also, the assay required small sample size and simple protein and phospholipid removal cartridge extraction without additional procedures. The sensitive method was successfully applied for the detection of xylazine and 2,6-dimethylaniline in a blood sample for clinical diagnosis and forensic toxicology.

Competing Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.