Research Article

Development and Validation of HLPC Method for the Estimation of Lamotrigine in Bulk and Pharmaceutical Formulations

Table 1

Precision of the method.

Intraday precisionInterday precision
InjectionAreaInjectionArea

Injection-13108761Injection-13139181
Injection-23113250Injection-23126470
Injection-33123530Injection-33135139
Injection-43129896Injection-43143320
Injection-53136552Injection-53143734

Average3122398Average3137569
Standard deviation11483.6Standard deviation7120.1
% RSD0.37% RSD0.23

Acceptance criteria:   the % RSD should be between less than 2.0.